array:20 [
  "pii" => "13084328"
  "issn" => "03002896"
  "doi" => "10.1016/S0210-5705(09)71003-9"
  "estado" => "S300"
  "fechaPublicacion" => "2005-12-15"
  "documento" => "article"
  "crossmark" => 0
  "subdocumento" => "fla"
  "cita" => "Arch Bronconeumol. 2005;41 Supl 4:68-74"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 50654
    "formatos" => array:3 [
      "EPUB" => 133
      "HTML" => 48099
      "PDF" => 2422
    ]
  ]
  "itemSiguiente" => array:16 [
    "pii" => "13084329"
    "issn" => "03002896"
    "doi" => "10.1016/S0210-5705(09)71003-9"
    "estado" => "S300"
    "fechaPublicacion" => "2005-12-15"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Arch Bronconeumol. 2005;41 Supl 4:75-80"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 14826
      "formatos" => array:3 [
        "EPUB" => 111
        "HTML" => 13717
        "PDF" => 998
      ]
    ]
    "es" => array:7 [
      "idiomaDefecto" => true
      "titulo" => "Tratamiento quirúrgico del SAHS y la roncopatía crónica"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "75"
          "paginaFinal" => "80"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13084329?idApp=UINPBA00003Z"
    "url" => "/03002896/00000041000000S4/v0_201307090955/13084329/v0_201307090956/es/main.assets"
  ]
  "itemAnterior" => array:16 [
    "pii" => "13084326"
    "issn" => "03002896"
    "doi" => "10.1016/S0210-5705(09)71003-9"
    "estado" => "S300"
    "fechaPublicacion" => "2005-12-15"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Arch Bronconeumol. 2005;41 Supl 4:51-67"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 74089
      "formatos" => array:3 [
        "EPUB" => 113
        "HTML" => 71474
        "PDF" => 2502
      ]
    ]
    "es" => array:7 [
      "idiomaDefecto" => true
      "titulo" => "Tratamiento del SAHS con presión continua positiva en la vía respiratoria superior (CPAP)"
      "tienePdf" => "es"
      "tieneTextoCompleto" => "es"
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "51"
          "paginaFinal" => "67"
        ]
      ]
      "contieneTextoCompleto" => array:1 [
        "es" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13084326?idApp=UINPBA00003Z"
    "url" => "/03002896/00000041000000S4/v0_201307090955/13084326/v0_201307090956/es/main.assets"
  ]
  "es" => array:8 [
    "idiomaDefecto" => true
    "titulo" => "Dispositivos de avance mandibular (DAM) en el tratamiento del SAHS"
    "tieneTextoCompleto" => true
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "68"
        "paginaFinal" => "74"
      ]
    ]
    "textoCompleto" => "<p class="elsevierStylePara">La utilizaci&#243;n de los aparatos intraorales en el tratamiento de los problemas obstructivos de la V&#237;a a&#233;rea superior &#40;VAS&#41; no es un concepto nuevo&#46; Ya en 1902&#44; Pierre Robin preconizaba la utilizaci&#243;n de un aparato de estas mismas caracter&#237;sticas &#40;monobloc&#41; con el objeto de realizar un adelantamiento funcional mandibular&#44; llevando a esta &#250;ltima hasta una posici&#243;n m&#225;s avanzada&#46; Con ello se consegu&#237;a un arrastre a&#241;adido de la lengua y se evitaba la ca&#237;da lingual hacia atr&#225;s &#40;glosoptosis&#41; que aparec&#237;a durante el dec&#250;bito supino en ni&#241;os con hipoplasia mandibular<span class="elsevierStyleSup">1</span>&#46; Las primeras publicaciones de la aparatolog&#237;a intraoral en relaci&#243;n con el SAHS aparecen en la d&#233;cada de los a&#241;os ochenta del pasado siglo<span class="elsevierStyleSup">2-5</span>&#44; como un intento de buscar tratamientos alternativos&#44; tanto a los procedimientos quir&#250;rgicos como a la presi&#243;n positiva continua en la v&#237;a respiratoria superior &#40;CPAP&#41;&#46;</p><p class="elsevierStylePara">Aunque se han descrito en el mercado m&#225;s de 50 tipos de aparatos<span class="elsevierStyleSup">6&#44;7</span> eficaces para el tratamiento del ronquido&#44; los aparatos de avance mandibular &#40;DAM&#41; en sus dos versiones &#40;avance fijo y avance regulable&#41; son los realmente eficaces en el manejo de los problemas obstructivos de la VAS&#46;</p><p class="elsevierStylePara">La American Sleep Disorders Association &#40;ASDA&#41; define los DAM<span class="elsevierStyleSup">8</span> como dispositivos que se introducen en la boca y modifican la posici&#243;n de la mand&#237;bula&#44; la lengua y otras estructuras de soporte de la VAS para el tratamiento del ronquido y&#47;o el SAHS&#46; Los considera como una alternativa v&#225;lida de primera elecci&#243;n para roncadores simples&#44; pacientes SAHS leve&#44; SAHS leve-moderado con bajo &#237;ndice de masa corporal&#44; pacientes con s&#237;ndrome de resistencia aumentada de la v&#237;a respiratoria superior &#40;SARVRS&#41; y como segunda elecci&#243;n en pacientes que no responden o rechazan los aparatos de presi&#243;n positiva&#44; pacientes con riesgo quir&#250;rgico elevado y con deficiente respuesta al tratamiento quir&#250;rgico&#46;</p><p class="elsevierStylePara">Los DAM realizan un movimiento anterior e inferior de la mand&#237;bula generando variaciones anat&#243;micas en la VAS que consiguen incrementar el &#225;rea seccional far&#237;ngea&#46; Este movimiento estabiliza y fija la mand&#237;bula y el hueso hioides&#44; lo que impide la posterorrotaci&#243;n de estas estructuras durante el dec&#250;bito y evita la ocupaci&#243;n de la v&#237;a respiratoria&#46;</p><p class="elsevierStylePara">Se ha comprobado que en los individuos SAHS el porcentaje total de tiempo en el que los pacientes duermen con la boca abierta m&#225;s de 5 mm &#40;rotaci&#243;n posterior y retrusi&#243;n&#41; es significativamente mayor &#40;69&#44;3&#37;&#41; que en individuos sanos &#40;11&#44;1&#37;&#41;&#59; por tanto&#44; el movimiento de rotaci&#243;n y fijaci&#243;n de la mand&#237;bula durante el sue&#241;o puede ser uno de los elementos importantes durante el tratamiento con DAM&#46; Este concepto puede explicar&#44; adem&#225;s&#44; por qu&#233; algunos pacientes obtienen buena respuesta con avances peque&#241;os incluso sin que se hayan evidenciado cambios en las VAS durante los estudios de la luz far&#237;ngea en vigilia<span class="elsevierStyleSup">9&#44;10</span>&#46; Aunque su mayor efecto aparece en la zona velofar&#237;ngea&#44; tienen repercusiones en todos los segmentos far&#237;ngeos&#46; Inicialmente aparece un aumento de rigidez y un ensanchamiento del espacio entre los pilares anterior y posterior de la faringe&#46; Se ha podido corroborar tambi&#233;n con resonancia magn&#233;tica<span class="elsevierStyleSup">11</span> que se produce una reducci&#243;n significativa del espesor de las paredes laterales far&#237;ngeas que es superior incluso al incremento luminal anteroposterior<span class="elsevierStyleSup">12</span>&#46; El palatogloso y el palatofar&#237;ngeo se tensan&#44; con lo que se reduce o incluso cesa la vibraci&#243;n de los tejidos blandos<span class="elsevierStyleSup">13</span>&#46; El paladar blando se desplaza ventralmente y aumenta el calibre de las paredes laterales de la zona velofar&#237;ngea<span class="elsevierStyleSup">11&#44;14</span>&#44; hechos ambos que por s&#237; mismos reducen dr&#225;sticamente el ronquido<span class="elsevierStyleSup">15</span>&#46;</p><p class="elsevierStylePara">La rotaci&#243;n horaria de la mand&#237;bula y el aumento pasivo de la dimensi&#243;n vertical activan al m&#250;sculo geniogloso&#44; que ser&#225; en parte el causante de los cambios que acontecen en la lengua<span class="elsevierStyleSup">16</span>&#46; Lowe et al<span class="elsevierStyleSup">17</span> han comprobado electromiogr&#225;ficamente c&#243;mo se incrementa el tono de la musculatura lingual &#40;sobre todo el geniogloso&#41; tras insertar los DAM y c&#243;mo este vuelve a descender tras quitar el aparato&#46; La lengua se desplaza hacia delante y ocupa una posici&#243;n superior&#44; dificultando as&#237; su ca&#237;da hacia atr&#225;s durante el sue&#241;o<span class="elsevierStyleSup">18</span>&#46;</p><p class="elsevierStylePara">Los DAM generan cambios en las presiones far&#237;ngeas locales que contribuyen a normalizar las propiedades fisiol&#243;gicas de la VAS<span class="elsevierStyleSup">19</span>&#46; La relaci&#243;n anat&#243;mica alterada &#40;rotaci&#243;n y avance mandibular&#41; induce un aumento de la estimulaci&#243;n neurosensorial que incrementa el tono muscular motor y reduce la colapsabilidad de la VAS&#46; De la misma forma&#44; se incrementa el vector de acci&#243;n de los m&#250;sculos dilatadores far&#237;ngeos que&#44; al desplazarse centralmente&#44; aumentan su longitud y mejoran su contractilidad<span class="elsevierStyleSup">20-21</span>&#46; Los DAM realizan una funci&#243;n similar a la que se produce durante las maniobras de sedaci&#243;n consciente y de resucitaci&#243;n cardiopulmonar en las que&#44; tras asegurar y preservar la permeabilidad de la VAS liber&#225;ndola de cualquier objeto extra&#241;o&#44; se realiza una maniobra de tracci&#243;n anterior e inferior mandibular<span class="elsevierStyleSup">22</span>&#46;</p><p class="elsevierStylePara">El avance funcional mandibular induce cambios en la posici&#243;n del hueso hioides hacia una posici&#243;n m&#225;s adelantada&#46; Aparece ahora una nueva situaci&#243;n de equilibrio de la musculatura suprahioidea que favorecer&#237;a el aumento de volumen y la permeabilidad de la VAS<span class="elsevierStyleSup">23</span>&#46; Este incremento de las VAS ha sido documentado por numerosos autores con diferentes sistemas de exploraci&#243;n y t&#233;cnicas de imagen<span class="elsevierStyleSup">24-28</span>&#46;</p><p class="elsevierStylePara">Aunque la respuesta no es la misma en todos los pacientes&#44; algunos autores admiten que hay un efecto dependiente de la dosis<span class="elsevierStyleSup">29&#44;30</span>&#44; ya que al aumentar el grado de avance mandibular mejora la situaci&#243;n cl&#237;nica del paciente<span class="elsevierStyleSup">31-33</span>&#46; As&#237;&#44; conforme incrementamos la protrusi&#243;n aparece una mejor&#237;a tanto en el n&#250;mero como en la gravedad de las desaturaciones&#46;</p><p class="elsevierStylePara">En las &#250;ltimas d&#233;cadas se han utilizado t&#233;cnicas cefalom&#233;tricas &#40;CFM&#41; para la evaluaci&#243;n de la VAS y la severidad del SAHS&#46; Una de las mayores deficiencias de la CFM es que s&#243;lo permite evaluar la VAS en el plano anteroposterior y no permite identificar los cambios laterales&#46; Estos cambios s&#237; pueden cuantificarse con otros sistemas m&#225;s sofisticados de imagen&#44; como la tomograf&#237;a computarizada &#40;TC&#41; y la resonancia magn&#233;tica &#40;RM&#41; que&#44; aunque dan una informaci&#243;n m&#225;s cuantitativa de la VAS&#44; tienen el inconveniente de estar m&#225;s sujetos a error que la CFM&#46; As&#237;&#44; el corte tanto en la TC como en la RM debe de ser perfectamente perpendicular a la VAS&#44; puesto que si no se hace de esta forma estaremos midiendo &#225;reas de VAS oblicuas&#44; que ser&#225;n mayores al calibre luminal real&#46; Otra fuente de errores con estas t&#233;cnicas deriva de su naturaleza din&#225;mica&#59; es decir&#44; cuando el paciente deglute durante la prueba se reduce falsamente el calibre de la VAS y de nuevo puede dar lugar a interpretaciones err&#243;neas de estenosis de la VAS<span class="elsevierStyleSup">34&#44;36</span>&#46;</p><p class="elsevierStylePara">No obstante&#44; a pesar de las carencias se&#241;aladas los estudios cefalom&#233;tricos tienen una serie de ventajas<span class="elsevierStyleSup">37-39</span>&#44; ya que se aproximan m&#225;s a una apreciaci&#243;n objetiva de la anatom&#237;a real del individuo&#44; es una exploraci&#243;n no invasiva&#44; de bajo coste y pr&#225;cticamente no necesita colaboraci&#243;n por parte del paciente&#46; Hay numerosos estudios que han documentado diferencias entre los pacientes con y sin SAHS<span class="elsevierStyleSup">40-49</span>&#46; Entre estas diferencias se&#241;alamos algunas de las m&#225;s significativas&#58; mayor estrechamiento en el &#225;rea velofar&#237;ngea&#44; aumento de dimensiones del paladar blando&#44; distancia hioides-plano mandibular &#40;MPH&#41; aumentada&#44; incremento de los &#225;ngulos de compensaci&#243;n craneocervical<span class="elsevierStyleSup">48&#44;49</span>&#46; El an&#225;lisis CFM puede complicarse todav&#237;a m&#225;s a&#241;adiendo nuevas variables&#58; edad&#44; sexo&#44; biotipo facial&#44; fenotipo&#44; posici&#243;n del paciente &#40;dec&#250;bito o bipedestaci&#243;n&#41;&#46;</p><p class="elsevierStylePara">Otros autores han relacionado algunas caracter&#237;sticas CFM con el fracaso de los DAM<span class="elsevierStyleSup">44</span> e incluso con el empeoramiento de la sintomatolog&#237;a que aparece en algunos pacientes tras su colocaci&#243;n<span class="elsevierStyleSup">50</span>&#46; Esto implica que algunas variaciones esquel&#233;ticas&#44; como las que cursan con entrecruzamiento dentario excesivo &#40;sobremordida&#41;&#44; pueden llegar a tener dos efectos adversos&#46; Por una parte disminuyen la eficacia del DAM&#44; puesto que se necesita una mayor apertura para conseguir un m&#237;nimo avance&#59; en segundo lugar&#44; la posterorrotaci&#243;n necesaria para vencer la sobremordida lleva hacia atr&#225;s al hioides y al resto de las estructuras&#44; favoreciendo la reducci&#243;n de la luz far&#237;ngea<span class="elsevierStyleSup">50-52</span>&#46;</p><p class="elsevierStylePara">Tangugsorn et al<span class="elsevierStyleSup">48&#44;53</span> y Miyamoto et al<span class="elsevierStyleSup">54</span> han documentado cambios en la postura de la lengua en posici&#243;n supina y en bipedestaci&#243;n&#46; En pacientes no apneicos&#44; las posiciones linguales cambian&#44; apareciendo un movimiento dorsal de la lengua hacia la luz far&#237;ngea&#46; Este hecho no aparece por el contrario en pacientes SAHS&#44; donde el cambio posicional &#40;durante la vigilia&#41; no implica variaciones en la posici&#243;n lingual&#44; como un mecanismo protector del colapso a&#233;reo secundario al cambio gravitacional&#46; Esta compensaci&#243;n durante la vigilia est&#225; producida por el incremento del tono muscular que los pacientes apneicos tienen respecto de pacientes no SAHS<span class="elsevierStyleSup">55&#44;56</span> y que desaparece durante el sue&#241;o favoreciendo la obstrucci&#243;n de la VAS<span class="elsevierStyleSup">57&#44;58</span>&#46;</p><p class="elsevierStylePara">Tambi&#233;n se ha asociado una serie de condiciones esquel&#233;ticas y faciales con una respuesta m&#225;s favorable a la terapia con DAM como son&#58; altura facial anterior reducida&#44; base craneal anterior larga&#44; longitud maxilar aumentada&#44; plano mandibular reducido<span class="elsevierStyleSup">59</span>&#44; distancia MPH reducida y paladar blando corto<span class="elsevierStyleSup">60</span>&#44; ausencia de trastornos nasales&#44; grados I y II de Mallampati&#44; sujetos capaces de realizar una protrusi&#243;n mandibular &#62; 5-7 mm<span class="elsevierStyleSup">61</span>&#44; proporciones normales entre di&#225;metro de la v&#237;a a&#233;rea-paladar blando-lengua<span class="elsevierStyleSup">62</span>&#46; De la misma forma se ha asociado una mejor respuesta a los DAM en pacientes con bajo &#237;ndice de masa corporal<span class="elsevierStyleSup">63</span>&#44; circunferencia del cuello reducida&#44; pacientes j&#243;venes<span class="elsevierStyleSup">40</span>&#44; &#237;ndice de apnea-hipopnea &#40;IAH&#41; basal bajo y en episodios apneicos asociados a la posici&#243;n supina<span class="elsevierStyleSup">64</span>&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">Eficacia de los DAM</span></p><p class="elsevierStylePara">Inicialmente se comenz&#243; con la publicaci&#243;n de <span class="elsevierStyleItalic">case reports</span>&#44; luego aparecieron series de casos cada vez m&#225;s numerosos y ahora&#44; tras varios ensayos cl&#237;nicos bien controlados<span class="elsevierStyleSup">13&#44;18&#44;33&#44;65</span>&#44; podemos afirmar que&#44; a pesar de la considerable variabilidad de dise&#241;os que encontramos entre los DAM&#44; hay suficientes avales en la bibliograf&#237;a cient&#237;fica que demuestran su eficacia<span class="elsevierStyleSup">18&#44;64-66</span>&#46; Todos estos estudios publicados documentan polisomnogr&#225;ficamente la validez del m&#233;todo&#46; Aunque en un principio el dise&#241;o del aparato no parec&#237;a tener relaci&#243;n con su efectividad<span class="elsevierStyleSup">67</span>&#44; recientemente se ha descrito que incluso se puede mejorar el porcentaje de &#233;xitos con el uso de los DAM con avance regulable<span class="elsevierStyleSup">29&#44;30&#44;33</span>&#44; con lo que se concluye que cuanto m&#225;s agresivos resultan los protocolos&#44; tanto mayor es la relaci&#243;n de &#233;xitos<span class="elsevierStyleSup">18&#44;68</span>&#46; En este mismo orden de cosas han aparecido algunos estudios<span class="elsevierStyleSup">69-72</span> en los que&#44; al igual que se realiza con la CPAP <span class="elsevierStyleItalic">&#40;single-nigth titration&#41;</span>&#44; se ha ensayado la titulaci&#243;n de los DAM de avance regulable durante la noche del estudio&#46; Con ello se conseguir&#237;a la dosis &#243;ptima terap&#233;utica y se acortar&#237;a de manera notable el protocolo de avance&#46; Estos ensayos&#44; aunque prometedores&#44; resultan dif&#237;ciles de estandarizar dada la alta variabilidad de tolerancia y de respuesta individual al tratamiento&#46;</p><p class="elsevierStylePara">Hay tambi&#233;n pacientes en los que no se consiguen buenos resultados&#58; nos referimos a algunos SAHS severos&#44; ya que aunque hay evidencias y datos positivos al respecto&#44; el porcentaje de &#233;xitos resulta inferior en relaci&#243;n con los leves y moderados<span class="elsevierStyleSup">68&#44;73&#44;74</span>&#44; pacientes con protrusiva deficiente &#40;escaso avance mandibular&#41;&#44; gran sobremordida&#44; casos con problemas dentales y&#47;o periodontales y pacientes con patolog&#237;a de la articulaci&#243;n temporomandibular &#40;ATM&#41;&#46;</p><p class="elsevierStylePara">Los DAM son muy eficaces en el tratamiento del ronquido &#40;eliminaci&#243;n en el 50&#37; de los casos&#41;<span class="elsevierStyleSup">67&#44;73-75</span> y consiguen una reducci&#243;n significativa del mismo en 90&#37;-100&#37; de los pacientes<span class="elsevierStyleSup">65&#44;73&#44;74</span>&#44; as&#237; como una mejora en la calidad del sue&#241;o<span class="elsevierStyleSup">67&#44;76&#44;77</span>&#46; El ronquido se asocia tambi&#233;n a una entidad intermedia&#44; como es el SARVAS&#44; en el que aparece somnolencia junto con un aumento anormal de la presi&#243;n esof&#225;gica&#44; aunque sin apneas&#47;hipopneas mensurables<span class="elsevierStyleSup">78</span>&#46; La remisi&#243;n o reducci&#243;n del ronquido en estos casos sugiere que los DAM pueden ser el tratamiento de elecci&#243;n<span class="elsevierStyleSup">79</span>&#46;</p><p class="elsevierStylePara">En los registros polisomnogr&#225;ficos se confirman los beneficios percibidos por el paciente&#44; con reducci&#243;n de la frecuencia e intensidad del ronquido&#44; apneas&#44; n&#250;mero de <span class="elsevierStyleItalic"> arousals</span> y un aumento significativo del sue&#241;o de ondas lentas y REM<span class="elsevierStyleSup">18&#44;64&#44;80&#44;81</span>&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">DAM frente a CPAP</span></p><p class="elsevierStylePara">La CPAP es capaz de controlar la mayor&#237;a de los casos&#44; mientras que los DAM no poseen esta eficacia&#46; Esta ventaja de base de la CPAP est&#225; compensada&#44; en parte&#44; por la mejor aceptaci&#243;n de los DAM por los pacientes&#46; Incluso los individuos con buena respuesta a ambas terapias suelen decantarse hacia los DAM en un porcentaje muy significativo<span class="elsevierStyleSup">18&#44;33&#44;82</span>&#46; Ferguson et al<span class="elsevierStyleSup">33</span> compararon la eficacia&#44; los efectos secundarios&#44; el seguimiento de tratamiento y las preferencias individuales en un estudio prospectivo aleatorizado&#46; El IAH era inferior con CPAP &#40;3&#44;5 &#177; 1&#44;6&#41; que con DAM &#40;9&#44;7 &#177; 7&#44;3&#41;&#46; Otros estudios<span class="elsevierStyleSup">18&#44;74</span> muestran cifras similares en cuanto a la saturaci&#243;n de ox&#237;geno&#44; que tambi&#233;n es superior con la CPAP&#46; Pero al analizar la satisfacci&#243;n personal&#44; el seguimiento y los efectos secundarios&#44; los DAM fueron superiores a la CPAP&#46; La preferencia de los DAM frente a otras terapias se explica por diferentes razones&#44; entre las que destacan la comodidad&#44; el ruido&#44; la necesidad de abastecimiento el&#233;ctrico&#44; la claustrofobia y la irritaci&#243;n tanto cut&#225;nea como de la VAS&#46; Se ha publicado que el uso de los DAM es menos efectivo que la CPAP en pacientes con SAHS severos y que debe limitarse a casos leves y moderados<span class="elsevierStyleSup">18&#44;64&#44;82</span>&#46; Otros estudios&#44; sin embargo&#44; no encuentran razones para excluir a algunos de estos pacientes y abogan por su utilizaci&#243;n incluso en casos severos&#44; con buenos resultados aunque con gran variabilidad individual<span class="elsevierStyleSup">65&#44;75&#44;76&#44;83</span>&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold"> Uvulopalatofaringoplastia frente a DAM</span></p><p class="elsevierStylePara">En relaci&#243;n con algunos procedimientos quir&#250;rgicos&#44; como la uvulopalatofaringoplastia&#44; hay trabajos donde los DAM se confirman superiores en cuanto a preferencia individual y eficacia de tratamiento<span class="elsevierStyleSup">84&#44;85</span>&#46; Pero adem&#225;s&#44; incluso en situaciones de fracasos quir&#250;rgicos<span class="elsevierStyleSup">86</span>&#44; los DAM han sido utilizados con &#233;xito como &#34;tratamiento de rescate&#34; y han mejorado la sintomatolog&#237;a en m&#225;s del 50&#37; de los pacientes<span class="elsevierStyleSup">84&#44;87</span>&#46; Algunos aparatos intraorales&#44; como los elevadores del paladar &#40;Palatal Lifting Devices&#41;&#44; a pesar de que son poco utilizados por sus escasos resultados y la escasa tolerancia por parte de los pacientes&#44; pueden utilizarse como factores <span class="elsevierStyleItalic">stent</span> tras una uvulopalatofaringoplastia&#44; ya que dificultan la retracci&#243;n cicatrizal generada por &#233;sta<span class="elsevierStyleSup">88</span>&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleBold">Protocolos de actuaci&#243;n</span></p><p class="elsevierStylePara">Al comparar los riesgos y los beneficios del tratamiento con DAM respecto a las otras terapias disponibles podemos sugerir que los DAM representan una alternativa &#250;til para estos pacientes&#46; Los criterios ASDA que antes hemos mencionado suponen un buen punto de partida&#44; pero tambi&#233;n hay otros factores que pueden ayudar al cl&#237;nico de la unidad del sue&#241;o a remitir al paciente a un profesional cualificado en el manejo de los DAM&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleItalic">S&#237; DAM</span></p><p class="elsevierStylePara">&#173; Ronquido&#58; s&#237;ntoma principal&#46;</p><p class="elsevierStylePara">&#173; Rechazo&#47;intolerancia a la CPAP&#46;</p><p class="elsevierStylePara">&#173; Fallo quir&#250;rgico&#46;</p><p class="elsevierStylePara">&#173; Viajes frecuentes&#46;</p><p class="elsevierStylePara">&#173; Claustrofobia y&#47;o problemas de ajuste con la mascarilla&#46;</p><p class="elsevierStylePara"><span class="elsevierStyleItalic">No DAM</span></p><p class="elsevierStylePara">&#173; Somnolencia&#58; s&#237;ntoma principal&#46;</p><p class="elsevierStylePara">&#173; Dentici&#243;n insuficiente para dar estabilidad al DAM&#46;</p><p class="elsevierStylePara">&#173; Enfermedad activa de ATM&#46;</p><p class="elsevierStylePara">&#173; Facilidad para las na&#250;seas&#46;</p><p class="elsevierStylePara">&#173; Obesidad m&#243;rbida&#46;</p><p class="elsevierStylePara">&#173; Desaturaci&#243;n de O<span class="elsevierStyleInf">2</span> severa&#46;</p><p class="elsevierStylePara">Si el ronquido es el s&#237;ntoma principal&#44; hay poca sintomatolog&#237;a diurna &#40;somnolencia&#41; y la hipoxemia no es un factor preponderante&#44; los DAM pueden ser el tratamiento de elecci&#243;n&#46; Pero cuando la somnolencia es el principal s&#237;ntoma&#44; la CPAP supone una mejor elecci&#243;n&#44; puesto que la respuesta positiva al tratamiento mejora y asegura la aceptaci&#243;n de &#233;ste<span class="elsevierStyleSup">18&#44;33</span>&#46; Los pacientes dependientes de CPAP pueden utilizar tambi&#233;n los DAM como coadyuvantes del tratamiento&#44; sustituyendo al aparato de presi&#243;n positiva ante situaciones especiales&#44; como puede ser la realizaci&#243;n de un viaje&#46; Los casos severos con hipoxemia y somnolencia diurna&#44; por el contrario&#44; pueden estar infratratados si s&#243;lo se utilizan los DAM y estos pacientes se adaptan mejor a la CPAP o a la terapia combinada&#46;</p><p class="elsevierStylePara">La continuidad de tratamiento con DAM tampoco es perfecta y la mayor&#237;a de los datos publicados corresponden a pacientes con buenos resultados&#46; No obstante&#44; la experiencia con CPAP<span class="elsevierStyleSup">89</span> sugiere que los cuestionarios realizados a los pacientes quiz&#225; tiendan a sobrestimar su actual utilizaci&#243;n<span class="elsevierStyleSup">90</span>&#46; Fergusson et al<span class="elsevierStyleSup">18&#44;33</span> detallan hasta un 24&#37; de incumplimiento en pacientes con DAM sin avance y s&#243;lo un 5&#37; en los que utilizan DAM regulable&#46; Otros autores objetivan que el seguimiento a largo plazo suele ser bastante regular&#44; con una media de utilizaci&#243;n nocturna de 6&#44;8 h&#47;noche<span class="elsevierStyleSup">68&#44;91</span> &#40;rango&#44; 5&#44;6-3&#44;5 h&#41;&#46;</p><p class="elsevierStylePara">La ca&#237;da o expulsi&#243;n del aparato durante la noche era algo frecuente hace unos a&#241;os pero ahora&#44; con los nuevos dise&#241;os y los nuevos materiales que se han incorporado&#44; ha quedado reducida a alguna an&#233;cdota aislada&#46;</p><p class="elsevierStylePara">Los efectos secundarios pueden diferenciarse en tres tipos&#58; cefalom&#233;tricos&#44; entre los que aparecen peque&#241;os movimientos mandibulares &#40;avance 0&#44;1 mm&#59; descenso 0&#44;3 mm&#41; rotaci&#243;n mandibular posterior &#40;0&#44;5&#186;&#41;&#44; aumento de la longitud mandibular &#40;0&#44;4 mm&#41; aunque este &#250;ltimo dato corresponde a los casos con grandes avances&#46; En segundo lugar aparecen peque&#241;os cambios oclusales con reducci&#243;n de resalte &#40;0&#44;2-1&#44;3 mm&#41; y sobremordida &#40;0&#44;1-1&#44;8 mm&#41;&#44; vestibuloversi&#243;n incisiva superior &#40;2&#186;&#41;&#44; linguoversi&#243;n incisiva inferior &#40;4&#44;5&#186;&#41; y movimiento mesial de los molares inferiores &#40;0&#44;4 mm&#41;&#46; En tercer lugar se describen las molestias universales ante el inicio de cualquier tratamiento ortod&#243;ncico y que van disminuyendo con el tiempo de uso del aparato&#58; exceso de salivaci&#243;n&#44; sequedad de mucosas&#44; sensibilidad dental&#44; molestias articulares &#40;ATM&#41; y musculares &#40;catalogadas como dolor miofascial que aparece al quitar el DAM y persisten durante un tiempo&#41;&#46; La mayor&#237;a de autores ha encontrado una mayor incidencia de estos efectos en los DAM de avance fijo respecto de los de avance regulable&#46; Con los protocolos m&#225;s agresivos de avance quiz&#225; deber&#237;a esperarse una mayor incidencia de estas complicaciones&#44; pero hasta el momento esto no ha sido as&#237;<span class="elsevierStyleSup">76-77&#44;92-96</span>&#46;</p><p class="elsevierStylePara">La principal raz&#243;n para dejar el tratamiento es la p&#233;rdida de eficacia&#59; sin embargo&#44; este hecho quiz&#225; pueda paliarse con la ayuda de los dispositivos de avance regulable&#44; aunque con ambos se han documentado &#233;xitos terap&#233;uticos incluso en obstrucciones far&#237;ngeas m&#225;s bajas<span class="elsevierStyleSup">97</span>&#46; En cuanto a las contraindicaciones absolutas para la utilizaci&#243;n de los DAM&#44; podemos encontrar el insuficiente n&#250;mero de dientes para soportar&#47;retener el aparato&#44; la presencia de una enfermedad activa de la articulaci&#243;n temporomandibular y&#44; finalmente&#44; la movilidad dentaria grave y&#47;o la enfermedad periodontal no tratada<span class="elsevierStyleSup">98</span>&#46; Pero incluso en pacientes en los que concurran dos de las contraindicaciones absolutas para los DAM &#40;ed&#233;ntulos totales y enfermedad de la articulaci&#243;n temporomandibular&#41;&#44; todav&#237;a queda un tipo de aparatolog&#237;a intraoral que se debe tener en cuenta&#58; los aparatos retenedores de la lengua &#40;TRD&#41; que&#44; adem&#225;s&#44; pueden utilizarse con &#233;xito en los pacientes que presentan macroglosia franca y en los que por diferentes razones no pueden realizar un avance mandibular adecuado<span class="elsevierStyleSup">2&#44;99-102</span>&#46;</p><p class="elsevierStylePara">El manejo de los DAM requiere formaci&#243;n dental cualificada&#44; tanto para el dise&#241;o como para su ajuste y mantenimiento&#46; La fabricaci&#243;n del aparato supone s&#243;lo el primer eslab&#243;n del tratamiento&#46; Los DAM deben de acomodarse seg&#250;n unos protocolos para alcanzar resultados &#243;ptimos&#44; tanto de eficacia como de comodidad para los usuarios&#46; Este proceso puede llevar semanas o incluso meses y debe de evaluarse juntamente con las unidades del sue&#241;o&#46; La European Dental Sleep Medicine Academy<span class="elsevierStyleSup">103</span> ha realizado un protocolo en detalle para la utilizaci&#243;n de estos aparatos&#46; Este mismo organismo recomienda que antes de colocar un DAM en un paciente SAHS se realicen dos cosas&#58; la primera&#44; un diagn&#243;stico y una prescripci&#243;n del aparato en regla enviado al dentista con formaci&#243;n especializada en este tipo de aparatos&#44; por parte de los m&#233;dicos responsables de las unidades del sue&#241;o&#44; y la segunda&#44; un examen oral adecuado realizado por el profesional encargado del dise&#241;o y colocaci&#243;n del DAM &#40;historia m&#233;dica y dental&#44; valoraci&#243;n de tejidos blandos&#44; evaluaci&#243;n periodontal&#44; examen oclusal y temporomandibular&#44; evaluaci&#243;n de h&#225;bitos y parafunciones&#44; examen detallado de los dientes y restauraciones presentes&#44; as&#237; como las radiograf&#237;as y modelos de escayola pertinentes&#41;&#46; Es el cl&#237;nico experimentado quien selecciona y dise&#241;a el aparato m&#225;s indicado para cada caso en particular&#46; Una vez que el aparato ha sido realizado y ajustado se entrena al paciente para su inserci&#243;n y los cuidados de mantenimiento&#46; Tras conseguir acostumbrar al paciente al aparato se comienza a realizar el avance gradualmente &#40;suele requerir semanas o meses de ajuste&#41; hasta que se consigue una posici&#243;n confortable y adecuada para aliviar el ronquido y&#47;o la apnea del sue&#241;o&#46; Tras esto el paciente debe ser revaluado en la unidad del sue&#241;o&#46; Ante cualquier cambio negativo ha de realizarse una nueva visita al laboratorio del sue&#241;o y realizar de nuevo el diagn&#243;stico&#46; En estas circunstancias puede llegar a ser preceptivo un posible redise&#241;o o incluso el cambio de aparato&#46; Los pacientes en tratamiento con DAM necesitan una media de visitas regulares durante los primeros 12 meses para asegurar el bienestar&#44; la retenci&#243;n y los resultados adecuados para la enfermedad&#46;</p><p class="elsevierStylePara">Los DAM son una alternativa razonable pues no suponen un cambio permanente en los individuos &#40;como ocurre con la cirug&#237;a&#41; y pueden desestimarse en cualquier momento&#46; El estudio adecuado del paciente en cuanto a evaluaci&#243;n de riesgos&#47;beneficios ha de exponerse individualmente<span class="elsevierStyleSup">104</span> para poder conseguir la mejor terapia en cada caso&#46;</p><p class="elsevierStylePara">Queda todav&#237;a un aspecto por plantear que debe tomarse en consideraci&#243;n&#46; Es la posible prevenci&#243;n de la enfermedad que nos ocupa en edades tempranas&#44; en las que con medios ortod&#243;ncicos es viable estimular o inhibir el crecimiento sagital y&#47;o transversal mandibular o maxilar y&#44; por ende&#44; inducir cambios en las VAS&#44; consiguiendo una conformaci&#243;n de la v&#237;a respiratoria menos susceptible al colapso<span class="elsevierStyleSup">105-107</span>&#46;</p>"
    "pdfFichero" => "6v41nSupl.4a13084328pdf001.pdf"
    "tienePdf" => true
    "bibliografia" => array:2 [
      "titulo" => "Bibliograf&#237;a"
      "seccion" => array:1 [
        0 => array:1 [
          "bibliografiaReferencia" => array:107 [
            0 => array:3 [
              "identificador" => "bib1"
              "etiqueta" => "1"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Glosoptosis due to atresia and hypotrophy of the mandible&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "Robin P&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Dis Child"
                        "fecha" => "1934"
                        "volumen" => "48"
                        "paginaInicial" => "541"
                        "paginaFinal" => "7"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            1 => array:3 [
              "identificador" => "bib2"
              "etiqueta" => "2"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The effects of a nonsurgical treatment for obstructive sleep apnea&#58; the tongue retaining device&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "Cartwright RD"
                            1 => "Samelson CF&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "J Am Med Assoc"
                        "fecha" => "1982"
                        "volumen" => "248"
                        "paginaInicial" => "705"
                        "paginaFinal" => "9"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            2 => array:3 [
              "identificador" => "bib3"
              "etiqueta" => "3"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Predicting response to the tongue retaining device for sleep apnea syndrome&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "Cartwright RD&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Arch Otolaryngol"
                        "fecha" => "1985"
                        "volumen" => "111"
                        "paginaInicial" => "385"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/4004636"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            3 => array:3 [
              "identificador" => "bib4"
              "etiqueta" => "4"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Treatment of obstructive sleep apnea with a nocturnal airway patency appliance&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "Soll BA"
                            1 => "George PT&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1056/NEJM198508083130612"
                      "Revista" => array:6 [
                        "tituloSerie" => "N Engl J Med"
                        "fecha" => "1985"
                        "volumen" => "313"
                        "paginaInicial" => "386"
                        "paginaFinal" => "7"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/4010757"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            4 => array:3 [
              "identificador" => "bib5"
              "etiqueta" => "5"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Sleep apnea syndrome&#46; Correction with surgical advancement of the mandible&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [
                            0 => "Bear SE"
                            1 => "Priest JH&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Oral Surg"
                        "fecha" => "1980"
                        "volumen" => "38"
                        "paginaInicial" => "543"
                        "paginaFinal" => "9"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/6929910"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            5 => array:3 [
              "identificador" => "bib6"
              "etiqueta" => "6"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The efficacy of a novel tongue stabilizing device on polisomnographic variables in sleep disordered breathing&#58; a pilot study&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-002-0069-1"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2002"
                        "volumen" => "6"
                        "paginaInicial" => "69"
                        "paginaFinal" => "76"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            6 => array:3 [
              "identificador" => "bib7"
              "etiqueta" => "7"
              "referencia" => array:1 [
                0 => array:2 [
                  "referenciaCompleta" => "Dental appliances for the treatment of snoring and obstructive sleep apnea. En: Kryger MH, Roth T, Dement WC, editors. Principles and practice of sleep medicine. Philadelphia: Saunders; 1994. p. 722-35."
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Dental appliances for the treatment of snoring and obstructive sleep apnea&#46; En&#58; Kryger MH&#44; Roth T&#44; Dement WC&#44; editors&#46; Principles and practice of sleep medicine&#46; Philadelphia&#58; Saunders&#59; 1994&#46; p&#46; 722-35&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [ …1]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            7 => array:3 [
              "identificador" => "bib8"
              "etiqueta" => "8"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Practice parameters for the treatment of snoring and obstructive sleep apnea with oral appliances&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [ …1]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "1995"
                        "volumen" => "18"
                        "paginaInicial" => "511"
                        "paginaFinal" => "3"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            8 => array:3 [
              "identificador" => "bib9"
              "etiqueta" => "9"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Mandibular posture during sleep in healthy adults&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Arch Oral Biol"
                        "fecha" => "1998"
                        "volumen" => "43"
                        "paginaInicial" => "269"
                        "paginaFinal" => "75"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            9 => array:3 [
              "identificador" => "bib10"
              "etiqueta" => "10"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Mandibular posture during sleep in patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Arch Oral Biol"
                        "fecha" => "1999"
                        "volumen" => "43"
                        "paginaInicial" => "657"
                        "paginaFinal" => "64"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            10 => array:3 [
              "identificador" => "bib11"
              "etiqueta" => "11"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Upper airway soft tissue structural changes with dental appliances in apneics&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:4 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "1996"
                        "volumen" => "153"
                        "paginaInicial" => "A719"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            11 => array:3 [
              "identificador" => "bib12"
              "etiqueta" => "12"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Mandibular advancement oral appliance therapy for obstructive sleep apnea&#58; effect on awake caliber of the velopharynx&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [ …5]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Thorax"
                        "fecha" => "1999"
                        "volumen" => "54"
                        "paginaInicial" => "9729"
                        "paginaFinal" => "77"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            12 => array:3 [
              "identificador" => "bib13"
              "etiqueta" => "13"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:1 [
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Efficacy and co-morbidity of oral appliances in the treatment of obstructive sleep apnea-hypopnea:a systematic review Clin Rev Oral Biol Med"
                        "fecha" => "2004"
                        "volumen" => "15"
                        "paginaInicial" => "137"
                        "paginaFinal" => "55"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            13 => array:3 [
              "identificador" => "bib14"
              "etiqueta" => "14"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Three-dimensional upper airway computedtomography in obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm/144.2.428"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am Rev Respir Dis"
                        "fecha" => "1991"
                        "volumen" => "144"
                        "paginaInicial" => "428"
                        "paginaFinal" => "32"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            14 => array:3 [
              "identificador" => "bib15"
              "etiqueta" => "15"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Ronchopatie chronique et syndrome de l&#39;apn&#233;e-hypopn&#233;e obstructive du sommeil chez l&#39;enfant&#46;"
                      "idioma" => "fr"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "L'orthodontie française"
                        "fecha" => "2003"
                        "volumen" => "74"
                        "paginaInicial" => "431"
                        "paginaFinal" => "57"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            15 => array:3 [
              "identificador" => "bib16"
              "etiqueta" => "16"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Three-dimensional CT reconstructions of tongue and airway in adult subjects with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "1986"
                        "volumen" => "90"
                        "paginaInicial" => "364"
                        "paginaFinal" => "74"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            16 => array:3 [
              "identificador" => "bib17"
              "etiqueta" => "17"
              "referencia" => array:1 [
                0 => array:2 [
                  "referenciaCompleta" => "Effects of mandibular repositioning appliance used in the treatment of obstructive sleep apnea on muscle activity. En: Suratt PM, Remmers JE, editors. Sleep and respiration. New York: Willey; 1990. p. 395-405."
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effects of mandibular repositioning appliance used in the treatment of obstructive sleep apnea on muscle activity&#46; En&#58; Suratt PM&#44; Remmers JE&#44; editors&#46; Sleep and respiration&#46; New York&#58; Willey&#59; 1990&#46; p&#46; 395-405&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            17 => array:3 [
              "identificador" => "bib18"
              "etiqueta" => "18"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A short term controlled trial of an adjustable oral appliance for the treatment of mild to moderate obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:6 [ …6]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Thorax"
                        "fecha" => "1997"
                        "volumen" => "52"
                        "paginaInicial" => "362"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            18 => array:3 [
              "identificador" => "bib19"
              "etiqueta" => "19"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effect of a prosthetic appliance for treatment of sleep apnea syndrome on masticatory and tongue muscle activity&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [ …1]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Prosthet Dent"
                        "fecha" => "1998"
                        "volumen" => "79"
                        "paginaInicial" => "537"
                        "paginaFinal" => "44"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            19 => array:3 [
              "identificador" => "bib20"
              "etiqueta" => "20"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Genioglossus muscle activity an inspiratory timing in obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "1993"
                        "volumen" => "104"
                        "paginaInicial" => "138"
                        "paginaFinal" => "45"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            20 => array:3 [
              "identificador" => "bib21"
              "etiqueta" => "21"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effects of a mandibular repositioning appliance used in the treatment of obstructive sleep apnea on tongue muscle activity&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Prog Clin Biol Res"
                        "fecha" => "1990"
                        "volumen" => "345"
                        "paginaInicial" => "395"
                        "paginaFinal" => "405"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            21 => array:3 [
              "identificador" => "bib22"
              "etiqueta" => "22"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effects of mandibular advancement on airway curvature and obstructive sleep apnea severity&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur Respir J"
                        "fecha" => "2004"
                        "volumen" => "23"
                        "paginaInicial" => "263"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            22 => array:3 [
              "identificador" => "bib23"
              "etiqueta" => "23"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Changes in airway and hyoid position in response to mandibular protrusion in subjects with obstructive sleep apnoea &#40;OSA&#41;&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:5 [ …5]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1999"
                        "volumen" => "2"
                        "paginaInicial" => "363"
                        "paginaFinal" => "76"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            23 => array:3 [
              "identificador" => "bib24"
              "etiqueta" => "24"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Three dimensional CT reconstruction of the tongue and airway in adults subjects with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "1986"
                        "volumen" => "90"
                        "paginaInicial" => "364"
                        "paginaFinal" => "74"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            24 => array:3 [
              "identificador" => "bib25"
              "etiqueta" => "25"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Changes in the upper airway of patients who wear a modified functional appliance to treat obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Int J Adult Orthod Orthognath Surg"
                        "fecha" => "1995"
                        "volumen" => "10"
                        "paginaInicial" => "53"
                        "paginaFinal" => "7"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            25 => array:3 [
              "identificador" => "bib26"
              "etiqueta" => "26"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Advancement of mandible improves velopharyngeal airway patency&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Appl Physiol"
                        "fecha" => "1995"
                        "volumen" => "79"
                        "paginaInicial" => "2132"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            26 => array:3 [
              "identificador" => "bib27"
              "etiqueta" => "27"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The role of lateral cephalometric radiography and fluoroscopy in assessing mandibular advancement in sleep-related disorders&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1998"
                        "volumen" => "20"
                        "paginaInicial" => "121"
                        "paginaFinal" => "32"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            27 => array:3 [
              "identificador" => "bib28"
              "etiqueta" => "28"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Do oral appliances enlarge the airway in patients with obstructive sleep apnoea&#63; A prospective computerized tomographic study&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "2000"
                        "volumen" => "22"
                        "paginaInicial" => "159"
                        "paginaFinal" => "68"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            28 => array:3 [
              "identificador" => "bib29"
              "etiqueta" => "29"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Dose-dependent effects of mandibular advancement of pharyngeal mechanics and nocturnal oxygenation in patients with sleep-disordered breathing&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "2000"
                        "volumen" => "117"
                        "paginaInicial" => "1065"
                        "paginaFinal" => "72"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            29 => array:3 [
              "identificador" => "bib30"
              "etiqueta" => "30"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A prospective randomized Study comparing two different degrees of mandibular advancement with a dental appliance in treatment of severe obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-003-0119-3"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2003"
                        "volumen" => "7"
                        "paginaInicial" => "119"
                        "paginaFinal" => "30"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            30 => array:3 [
              "identificador" => "bib31"
              "etiqueta" => "31"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effect of tirated mandibular advancement and jaw opening on the upper airway in nonapneic men&#58; a magnetic resonance imaging and cephalometric study&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "2004"
                        "volumen" => "125"
                        "paginaInicial" => "191"
                        "paginaFinal" => "9"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            31 => array:3 [
              "identificador" => "bib32"
              "etiqueta" => "32"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The effect of mandibular advancement device on apneas and sleep in patients with OSAS&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1998"
                        "volumen" => "113"
                        "paginaInicial" => "707"
                        "paginaFinal" => "13"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            32 => array:3 [
              "identificador" => "bib33"
              "etiqueta" => "33"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A randomized crossover study of an oral appliance vs nasal continuous positive airway pressure in the treatment of mild moderate obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1996"
                        "volumen" => "109"
                        "paginaInicial" => "1269"
                        "paginaFinal" => "75"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            33 => array:3 [
              "identificador" => "bib34"
              "etiqueta" => "34"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cephalometric and physiologic predictors of the efficacy of an adjustable oral appliance for treating obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "2001"
                        "volumen" => "120"
                        "paginaInicial" => "639"
                        "paginaFinal" => "47"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            34 => array:3 [
              "identificador" => "bib35"
              "etiqueta" => "35"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Magnetic resonance imaging of the upper airway in obstructive sleep apnea before and after oral appliance therapy&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chin J Dent Res"
                        "fecha" => "1999"
                        "volumen" => "2"
                        "paginaInicial" => "27"
                        "paginaFinal" => "35"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            35 => array:3 [
              "identificador" => "bib36"
              "etiqueta" => "36"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Upper airway distensibility and collapsability in patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1990"
                        "volumen" => "98"
                        "paginaInicial" => "84"
                        "paginaFinal" => "91"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            36 => array:3 [
              "identificador" => "bib37"
              "etiqueta" => "37"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The efficacy of a mandibular advancement splint in relation to cephalometric variables&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-002-0115-z"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2002"
                        "volumen" => "6"
                        "paginaInicial" => "115"
                        "paginaFinal" => "24"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            37 => array:3 [
              "identificador" => "bib38"
              "etiqueta" => "38"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cephalometric measurements in snorers&#44; non snorers and patients with sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Thorax"
                        "fecha" => "1991"
                        "volumen" => "46"
                        "paginaInicial" => "419"
                        "paginaFinal" => "23"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            38 => array:3 [
              "identificador" => "bib39"
              "etiqueta" => "39"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Habitual snoring with and without obstructive sleep apnea&#58; the importance of cephalometric variables&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Thorax"
                        "fecha" => "1992"
                        "volumen" => "47"
                        "paginaInicial" => "157"
                        "paginaFinal" => "61"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            39 => array:3 [
              "identificador" => "bib40"
              "etiqueta" => "40"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Obstructive sleep apnea and cephalometric roentgenograms&#46; The role of anatomic upper airway abnormalities in the definition of abnormal breathing during sleep&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1988"
                        "volumen" => "93"
                        "paginaInicial" => "1199"
                        "paginaFinal" => "205"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            40 => array:3 [
              "identificador" => "bib41"
              "etiqueta" => "41"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cephalometric analysis in patients with obstructive sleep apnea syndrome&#58; skleletal morpphology&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Laryngol Otol"
                        "fecha" => "1989"
                        "volumen" => "103"
                        "paginaInicial" => "287"
                        "paginaFinal" => "92"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            41 => array:3 [
              "identificador" => "bib42"
              "etiqueta" => "42"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Obstructive sleep apnea subtipes by cluster analysis&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofanc Orthop"
                        "fecha" => "1992"
                        "volumen" => "101"
                        "paginaInicial" => "542"
                        "paginaFinal" => "53"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            42 => array:3 [
              "identificador" => "bib43"
              "etiqueta" => "43"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "L&#39; inter&#234;t physiopathologique d&#39;une &#233;tude cephalometrique par t&#233;l&#233;radiographie et IRM dans le syndrome d&#39;apn&#233;e du sommeil&#44; deductions th&#233;rapeutiques&#46;"
                      "idioma" => "fr"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Ann Otolaryng (Paris)"
                        "fecha" => "1990"
                        "volumen" => "107"
                        "paginaInicial" => "156"
                        "paginaFinal" => "9"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            43 => array:3 [
              "identificador" => "bib44"
              "etiqueta" => "44"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cephalometric comparisons of craniofacial and upper airway structure by skeletal subtype and gender in patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "1996"
                        "volumen" => "110"
                        "paginaInicial" => "653"
                        "paginaFinal" => "64"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            44 => array:3 [
              "identificador" => "bib45"
              "etiqueta" => "45"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Relationship between body max index&#44; age and upper airway measurements in snorers and sleep apnea patients&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Eur J Respir"
                        "fecha" => "1996"
                        "volumen" => "9"
                        "paginaInicial" => "1801"
                        "paginaFinal" => "9"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            45 => array:3 [
              "identificador" => "bib46"
              "etiqueta" => "46"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The cephalometric morphology of patients with obstructive sleep apnea &#40;OSA&#41;&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1996"
                        "volumen" => "18"
                        "paginaInicial" => "557"
                        "paginaFinal" => "69"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            46 => array:3 [
              "identificador" => "bib47"
              "etiqueta" => "47"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Airway dimensions and head posture in obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1996"
                        "volumen" => "18"
                        "paginaInicial" => "571"
                        "paginaFinal" => "9"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            47 => array:3 [
              "identificador" => "bib48"
              "etiqueta" => "48"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Obstructive sleep apnea&#58; a cephalometric study&#46; Part I&#46; Cervico-craniofacial skeletal morphology&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1995"
                        "volumen" => "17"
                        "paginaInicial" => "45"
                        "paginaFinal" => "56"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            48 => array:3 [
              "identificador" => "bib49"
              "etiqueta" => "49"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Natural head posture&#44; upper airway morphology and obstructive sleep apnea severity in adults&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1998"
                        "volumen" => "20"
                        "paginaInicial" => "133"
                        "paginaFinal" => "43"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            49 => array:3 [
              "identificador" => "bib50"
              "etiqueta" => "50"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Relaci&#243;n entre la maloclusi&#243;n e incremento de la v&#237;a a&#233;rea utilizando un aparato intraoral en pacientes con s&#237;ndrome de apnea obstructiva del sue&#241;o&#46;"
                      "idioma" => "es"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "RCOE"
                        "fecha" => "2002"
                        "volumen" => "7"
                        "paginaInicial" => "359"
                        "paginaFinal" => "64"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            50 => array:3 [
              "identificador" => "bib51"
              "etiqueta" => "51"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The efficacy of a mandibular advancement splint in relation to cephalometric variables&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-002-0115-z"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2002"
                        "volumen" => "6"
                        "paginaInicial" => "115"
                        "paginaFinal" => "24"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            51 => array:3 [
              "identificador" => "bib52"
              "etiqueta" => "52"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effect of vertical dimension on efficacy of oral appliance therapy in obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/rccm.200204-342OC"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2002"
                        "volumen" => "166"
                        "paginaInicial" => "860"
                        "paginaFinal" => "4"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            52 => array:3 [
              "identificador" => "bib53"
              "etiqueta" => "53"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Obstructive sleep apnea&#58; a cephalometric study&#46; Part II&#46; Uvuloglossopharyngeal morphology&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1995"
                        "volumen" => "15"
                        "paginaInicial" => "57"
                        "paginaFinal" => "67"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            53 => array:3 [
              "identificador" => "bib54"
              "etiqueta" => "54"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effect on body position on tongue posture in awake patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Thorax"
                        "fecha" => "1997"
                        "volumen" => "52"
                        "paginaInicial" => "255"
                        "paginaFinal" => "9"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            54 => array:3 [
              "identificador" => "bib55"
              "etiqueta" => "55"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Influence of sleep onset on upper airway muscle activity in apnea patients versus normal controls&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm.153.6.8665050"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit CAre Med"
                        "fecha" => "1996"
                        "volumen" => "153"
                        "paginaInicial" => "1880"
                        "paginaFinal" => "7"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            55 => array:3 [
              "identificador" => "bib56"
              "etiqueta" => "56"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Compensatory hyperactivity of an upper airway dilator in bulldogs&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm/148.1.185"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am Rev Respir Dis"
                        "fecha" => "1993"
                        "volumen" => "148"
                        "paginaInicial" => "185"
                        "paginaFinal" => "94"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            56 => array:3 [
              "identificador" => "bib57"
              "etiqueta" => "57"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Electrical stimulation of the lingual musculature in obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "J Appl Physiol"
                        "fecha" => "1996"
                        "volumen" => "81"
                        "paginaInicial" => "643"
                        "paginaFinal" => "52"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            57 => array:3 [
              "identificador" => "bib58"
              "etiqueta" => "58"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Does upper airway muscle injury trigger a vicious cycle in obstructive sleep apnea&#63; A hypotesis&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "1996"
                        "volumen" => "19"
                        "paginaInicial" => "465"
                        "paginaFinal" => "71"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            58 => array:3 [
              "identificador" => "bib59"
              "etiqueta" => "59"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The efficacy of a mandibular advancement splint in relation to cephalometric variables&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-002-0115-z"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2002"
                        "volumen" => "6"
                        "paginaInicial" => "115"
                        "paginaFinal" => "24"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            59 => array:3 [
              "identificador" => "bib60"
              "etiqueta" => "60"
              "referencia" => array:1 [
                0 => array:2 [
                  "referenciaCompleta" => "Efficacy of a herbst mandibular advancement device in obstructive sleep apnea. Am J Crit Care Med. 1994149:905-9."
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Efficacy of a herbst mandibular advancement device in obstructive sleep apnea&#46; Am J Crit Care Med&#46; 1994149&#58;905-9&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            60 => array:3 [
              "identificador" => "bib61"
              "etiqueta" => "61"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cephalometric and physiologic predictors of the efficacy of an adjustable oral appliance for treating obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "2001"
                        "volumen" => "120"
                        "paginaInicial" => "639"
                        "paginaFinal" => "47"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            61 => array:3 [
              "identificador" => "bib62"
              "etiqueta" => "62"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Magnetic resonance imaging of the upper airway in obstructive sleep apnea before and after oral appliance theraphy&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chin J Dent Res"
                        "fecha" => "1999"
                        "volumen" => "2"
                        "paginaInicial" => "27"
                        "paginaFinal" => "35"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            62 => array:3 [
              "identificador" => "bib63"
              "etiqueta" => "63"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Cephalometric predictors for orthopaedic mandibular avancement in obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "1995"
                        "volumen" => "17"
                        "paginaInicial" => "35"
                        "paginaFinal" => "43"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            63 => array:3 [
              "identificador" => "bib64"
              "etiqueta" => "64"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The effect of mandibular advancement device on apneas and sleep in patients with OSAS&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1998"
                        "volumen" => "113"
                        "paginaInicial" => "707"
                        "paginaFinal" => "13"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            64 => array:3 [
              "identificador" => "bib65"
              "etiqueta" => "65"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A prospective randomized Study comparing two different degrees of mandibular advancement with a dental appliance in treatment of severe obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-003-0119-3"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2003"
                        "volumen" => "7"
                        "paginaInicial" => "119"
                        "paginaFinal" => "30"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            65 => array:3 [
              "identificador" => "bib66"
              "etiqueta" => "66"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Oral appliances for the treatment of snoring and obstructive sleep apnea&#58; a review&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "1995"
                        "volumen" => "18"
                        "paginaInicial" => "501"
                        "paginaFinal" => "10"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            66 => array:3 [
              "identificador" => "bib67"
              "etiqueta" => "67"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Oral appliances for the treatment of snoring and obstructive sleep apnea&#58; a review&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "1995"
                        "volumen" => "18"
                        "paginaInicial" => "501"
                        "paginaFinal" => "10"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            67 => array:3 [
              "identificador" => "bib68"
              "etiqueta" => "68"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The effects of klearway oral appliance on airway size and obstructive sep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "2000"
                        "volumen" => "23"
                        "paginaInicial" => "S172"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            68 => array:3 [
              "identificador" => "bib69"
              "etiqueta" => "69"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Oral appliance theraphy for obstructive sleep apnea syndrome&#58; progressive mandibular advancement during polisomnography&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Cranio"
                        "fecha" => "1998"
                        "volumen" => "16"
                        "paginaInicial" => "44"
                        "paginaFinal" => "50"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            69 => array:3 [
              "identificador" => "bib70"
              "etiqueta" => "70"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Remotely controlled mandibular positioner predicts efficacy of oral appliances in sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/rccm.200310-1446OC"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2004"
                        "volumen" => "170"
                        "paginaInicial" => "366"
                        "paginaFinal" => "70"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            70 => array:3 [
              "identificador" => "bib71"
              "etiqueta" => "71"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "One-nigth mandibular advancement titration for obstructive sleep apnea syndrome&#58; a pilot study&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm.165.8.2108056"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2002"
                        "volumen" => "165"
                        "paginaInicial" => "1150"
                        "paginaFinal" => "3"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            71 => array:3 [
              "identificador" => "bib72"
              "etiqueta" => "72"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Mandibular advancement titration for obstructive sleep apnea&#46; Optimization of the procedure by comparing clinical and oximetric parameters&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "2004"
                        "volumen" => "125"
                        "paginaInicial" => "1761"
                        "paginaFinal" => "7"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            72 => array:3 [
              "identificador" => "bib73"
              "etiqueta" => "73"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Treatment of snoring and obstructive sleep apnea with a dental orthosis&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1991"
                        "volumen" => "99"
                        "paginaInicial" => "1378"
                        "paginaFinal" => "85"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            73 => array:3 [
              "identificador" => "bib74"
              "etiqueta" => "74"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Mandibular advancement splint&#58; an appliance to treat snoring and obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm.151.1.7812552"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "1995"
                        "volumen" => "151"
                        "paginaInicial" => "194"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            74 => array:3 [
              "identificador" => "bib75"
              "etiqueta" => "75"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A cross over study comparing the efficacy of CPAP with MAD on patients with OSA&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1996"
                        "volumen" => "109"
                        "paginaInicial" => "1447"
                        "paginaFinal" => "83"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            75 => array:3 [
              "identificador" => "bib76"
              "etiqueta" => "76"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Oral appliance theraphy improves symptoms in obstructive sleep apnea&#58; a randomized controled trial&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:4 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2002"
                        "paginaInicial" => "166743"
                        "paginaFinal" => "8"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            76 => array:3 [
              "identificador" => "bib77"
              "etiqueta" => "77"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A randomized controled study of a mandibular advancement splint for obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Respir Cri Care Med"
                        "fecha" => "2001"
                        "volumen" => "163"
                        "paginaInicial" => "1457"
                        "paginaFinal" => "61"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            77 => array:3 [
              "identificador" => "bib78"
              "etiqueta" => "78"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A cause of excessive daytime sleepiness&#58; the upper airway resistance syndrome&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1993"
                        "volumen" => "104"
                        "paginaInicial" => "781"
                        "paginaFinal" => "7"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            78 => array:3 [
              "identificador" => "bib79"
              "etiqueta" => "79"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Successful treatment of upper airway resistance syndrome with an oral appliance&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Sleep Breathing"
                        "fecha" => "1997"
                        "volumen" => "2"
                        "paginaInicial" => "97"
                        "paginaFinal" => "101"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            79 => array:3 [
              "identificador" => "bib80"
              "etiqueta" => "80"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A randomized controled crossover tial of two oral appliances for sleep apnea treatment&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm.162.1.9908112"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2000"
                        "volumen" => "162"
                        "paginaInicial" => "246"
                        "paginaFinal" => "51"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            80 => array:3 [
              "identificador" => "bib81"
              "etiqueta" => "81"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The mandibular repositioning device&#58; role in the treatment of obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "1996"
                        "volumen" => "19"
                        "paginaInicial" => "794"
                        "paginaFinal" => "800"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            81 => array:3 [
              "identificador" => "bib82"
              "etiqueta" => "82"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "An oral elastic advancement device for obstructive sleep apnea&#46;"
                      "idioma" => "ro"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm.161.2.9903079"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2000"
                        "volumen" => "161"
                        "paginaInicial" => "420"
                        "paginaFinal" => "5"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            82 => array:3 [
              "identificador" => "bib83"
              "etiqueta" => "83"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Effects of a mandibular protruding device on the sleep of patients with obstructive sleep apnea and snoring problems&#46; A 2 year follow up&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-003-0131-7"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2003"
                        "volumen" => "7"
                        "paginaInicial" => "131"
                        "paginaFinal" => "41"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            83 => array:3 [
              "identificador" => "bib84"
              "etiqueta" => "84"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A prospective randomized study of dental appliance compared with UPPP in the treatment of OSA&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Acta Otolaryngol Scand"
                        "fecha" => "1999"
                        "volumen" => "119"
                        "paginaInicial" => "503"
                        "paginaFinal" => "9"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            84 => array:3 [
              "identificador" => "bib85"
              "etiqueta" => "85"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:1 [
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Laser assisted uvulopalatoplasty for snoring:does it meet the expectations? Eur Respir J"
                        "fecha" => "2004"
                        "volumen" => "24"
                        "paginaInicial" => "66"
                        "paginaFinal" => "70"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            85 => array:3 [
              "identificador" => "bib86"
              "etiqueta" => "86"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Laser-assisted UPPP for the management of OSA&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Arch Otolaryngol Head Neck Surg"
                        "fecha" => "2002"
                        "volumen" => "128"
                        "paginaInicial" => "429"
                        "paginaFinal" => "38"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            86 => array:3 [
              "identificador" => "bib87"
              "etiqueta" => "87"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The efficacy of oral appliances in the treatment of persistent sleep apnea after UPPP&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1998"
                        "volumen" => "113"
                        "paginaInicial" => "992"
                        "paginaFinal" => "6"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            87 => array:3 [
              "identificador" => "bib88"
              "etiqueta" => "88"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Resolution of severe sleep disordered breathing with a nasopharyngeal obturator in 2 cases of nasopharyngeal stenosis complicating uvulopalatopharyngoplasty&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-004-0049-8"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2004"
                        "volumen" => "8"
                        "paginaInicial" => "49"
                        "paginaFinal" => "55"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            88 => array:3 [
              "identificador" => "bib89"
              "etiqueta" => "89"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Claustrophobia and adherence to CPAP treatment&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1177/0193945904273283"
                      "Revista" => array:6 [
                        "tituloSerie" => "West J Nurs Res"
                        "fecha" => "2005"
                        "volumen" => "27"
                        "paginaInicial" => "307"
                        "paginaFinal" => "21"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            89 => array:3 [
              "identificador" => "bib90"
              "etiqueta" => "90"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Objective measurements of patterns of nasal CPAP use by patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm/147.4.887"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am Rev Respir Dis"
                        "fecha" => "1993"
                        "volumen" => "147"
                        "paginaInicial" => "887"
                        "paginaFinal" => "95"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            90 => array:3 [
              "identificador" => "bib91"
              "etiqueta" => "91"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Evaluation of variable mandibular advancement appliance for treatmet of snoring and sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Chest"
                        "fecha" => "1999"
                        "volumen" => "116"
                        "paginaInicial" => "1511"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            91 => array:3 [
              "identificador" => "bib92"
              "etiqueta" => "92"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Efectos adversos del uso de una f&#233;rula de avance mandibular en el tratamiento de los transtornos respiratorios del sue&#241;o&#46; Revisi&#243;n de la literatura&#46;"
                      "idioma" => "es"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Archivos de Odontoestomatología"
                        "fecha" => "2004"
                        "volumen" => "20"
                        "paginaInicial" => "88"
                        "paginaFinal" => "100"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            92 => array:3 [
              "identificador" => "bib93"
              "etiqueta" => "93"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Side effects of mandibular advancement devices for sleep apnea treatment&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/ajrccm.164.5.2003078"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Resp Crit Care Med"
                        "fecha" => "2001"
                        "volumen" => "164"
                        "paginaInicial" => "813"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            93 => array:3 [
              "identificador" => "bib94"
              "etiqueta" => "94"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Influence of mandibular protruding device on airway passages and dentofacial characteristics in obstructive apnea and snoring&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Dentofac Orthop"
                        "fecha" => "2002"
                        "volumen" => "122"
                        "paginaInicial" => "371"
                        "paginaFinal" => "9"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            94 => array:3 [
              "identificador" => "bib95"
              "etiqueta" => "95"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Craniomandibular status and function in patients with habitual snoring and obstructive sleep apnea after nocturnal treatment with a mandibular advancement splint&#58; a 2-year follow-up&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Eur J Orthod"
                        "fecha" => "2000"
                        "volumen" => "22"
                        "paginaInicial" => "53"
                        "paginaFinal" => "60"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            95 => array:3 [
              "identificador" => "bib96"
              "etiqueta" => "96"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Dental and skeletal changes after 4 years of obstructive sleep apnea treatment with a mandibular advancement device&#58; a prospective randomized study&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Dentofac Orthop"
                        "fecha" => "2003"
                        "volumen" => "124"
                        "paginaInicial" => "53"
                        "paginaFinal" => "60"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            96 => array:3 [
              "identificador" => "bib97"
              "etiqueta" => "97"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Multidisciplinary management of the airway in a trauma-induced brain injury patient&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-004-0165-5"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2004"
                        "volumen" => "8"
                        "paginaInicial" => "165"
                        "paginaFinal" => "70"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            97 => array:3 [
              "identificador" => "bib98"
              "etiqueta" => "98"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Mandibular advancement devices&#46; Rate of contraindications in 100 consecutive obstructive sleep apnea patients&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1164/rccm.2008167"
                      "Revista" => array:6 [
                        "tituloSerie" => "Am J Respir Crit Care Med"
                        "fecha" => "2002"
                        "volumen" => "166"
                        "paginaInicial" => "274"
                        "paginaFinal" => "8"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            98 => array:3 [
              "identificador" => "bib99"
              "etiqueta" => "99"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Toward a treatment logic for sleep apnea&#58;the place of tongue retaining device&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Behav Res Ther"
                        "fecha" => "1988"
                        "volumen" => "26"
                        "paginaInicial" => "121"
                        "paginaFinal" => "6"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            99 => array:3 [
              "identificador" => "bib100"
              "etiqueta" => "100"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A comparative study of treatments for positional sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "1991"
                        "volumen" => "14"
                        "paginaInicial" => "546"
                        "paginaFinal" => "52"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            100 => array:3 [
              "identificador" => "bib101"
              "etiqueta" => "101"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "The effect of tongue retaining device on awake genioglossus muscle activity in patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Am J Orthod Dentofac Orthop"
                        "fecha" => "1996"
                        "volumen" => "110"
                        "paginaInicial" => "28"
                        "paginaFinal" => "35"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            101 => array:3 [
              "identificador" => "bib102"
              "etiqueta" => "102"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "A tongue retaining device and sleep state genioglossus muscle activity in patients with obstructive sleep apnea&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [ …4]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1043/0003-3219(1996)066<0273:ATRDAS>2.3.CO;2"
                      "Revista" => array:6 [
                        "tituloSerie" => "Angle Orthod"
                        "fecha" => "1996"
                        "volumen" => "66"
                        "paginaInicial" => "273"
                        "paginaFinal" => "80"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            102 => array:3 [
              "identificador" => "bib103"
              "etiqueta" => "103"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Report from the first conference of European Dental Sleep Medicine Academy&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1007/s11325-005-0007-0"
                      "Revista" => array:6 [
                        "tituloSerie" => "Sleep Breath"
                        "fecha" => "2005"
                        "volumen" => "9"
                        "paginaInicial" => "41"
                        "paginaFinal" => "2"
                        "link" => array:1 [
                          0 => array:2 [ …2]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            103 => array:3 [
              "identificador" => "bib104"
              "etiqueta" => "104"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "Position statement of the American Academy of Sleep Medicine&#46;"
                      "idioma" => "en"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [ …1]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Sleep"
                        "fecha" => "2000"
                        "volumen" => "33"
                        "paginaInicial" => "1017"
                        "paginaFinal" => "8"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            104 => array:3 [
              "identificador" => "bib105"
              "etiqueta" => "105"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "El activador y las v&#237;as a&#233;reas superiores&#46;"
                      "idioma" => "es"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [ …1]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Rev Esp Orthod"
                        "fecha" => "1994"
                        "volumen" => "24"
                        "paginaInicial" => "311"
                        "paginaFinal" => "8"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            105 => array:3 [
              "identificador" => "bib106"
              "etiqueta" => "106"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:1 [
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "Ortodoncia y vías aéreas superiores RCOE"
                        "fecha" => "2002"
                        "volumen" => "7"
                        "paginaInicial" => "417"
                        "paginaFinal" => "27"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            106 => array:3 [
              "identificador" => "bib107"
              "etiqueta" => "107"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:3 [
                      "titulo" => "L&#39;Orthodontie et les voies a&#233;riennes sup&#233;rieures&#46;"
                      "idioma" => "fr"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [ …3]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:5 [
                        "tituloSerie" => "L'Orthodontie Française"
                        "fecha" => "2004"
                        "volumen" => "75"
                        "paginaInicial" => "31"
                        "paginaFinal" => "7"
                      ]
                    ]
                  ]
                ]
              ]
            ]
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/03002896/00000041000000S4/v0_201307090955/13084328/v0_201307090956/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "14589"
    "tipo" => "SECCION"
    "es" => array:2 [
      "titulo" => "Consenso Nacional sobre el s&#237;ndrome de apneas-hipopneas del sue&#241;o"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "es"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/03002896/00000041000000S4/v0_201307090955/13084328/v0_201307090956/es/6v41nSupl.4a13084328pdf001.pdf?idApp=UINPBA00003Z&text.app=https://archbronconeumol.org/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/13084328?idApp=UINPBA00003Z"
]
Share
Journal Information
Vol. 41. Issue S4.
Pages 68-74 (December 2005)
Share
Share
Download PDF
More article options
Vol. 41. Issue S4.
Pages 68-74 (December 2005)
Full text access
Dispositivos de avance mandibular (DAM) en el tratamiento del SAHS
Visits
76832
This item has received
Article information
Full Text
Bibliography
Download PDF
Statistics
Full Text

La utilización de los aparatos intraorales en el tratamiento de los problemas obstructivos de la Vía aérea superior (VAS) no es un concepto nuevo. Ya en 1902, Pierre Robin preconizaba la utilización de un aparato de estas mismas características (monobloc) con el objeto de realizar un adelantamiento funcional mandibular, llevando a esta última hasta una posición más avanzada. Con ello se conseguía un arrastre añadido de la lengua y se evitaba la caída lingual hacia atrás (glosoptosis) que aparecía durante el decúbito supino en niños con hipoplasia mandibular1. Las primeras publicaciones de la aparatología intraoral en relación con el SAHS aparecen en la década de los años ochenta del pasado siglo2-5, como un intento de buscar tratamientos alternativos, tanto a los procedimientos quirúrgicos como a la presión positiva continua en la vía respiratoria superior (CPAP).

Aunque se han descrito en el mercado más de 50 tipos de aparatos6,7 eficaces para el tratamiento del ronquido, los aparatos de avance mandibular (DAM) en sus dos versiones (avance fijo y avance regulable) son los realmente eficaces en el manejo de los problemas obstructivos de la VAS.

La American Sleep Disorders Association (ASDA) define los DAM8 como dispositivos que se introducen en la boca y modifican la posición de la mandíbula, la lengua y otras estructuras de soporte de la VAS para el tratamiento del ronquido y/o el SAHS. Los considera como una alternativa válida de primera elección para roncadores simples, pacientes SAHS leve, SAHS leve-moderado con bajo índice de masa corporal, pacientes con síndrome de resistencia aumentada de la vía respiratoria superior (SARVRS) y como segunda elección en pacientes que no responden o rechazan los aparatos de presión positiva, pacientes con riesgo quirúrgico elevado y con deficiente respuesta al tratamiento quirúrgico.

Los DAM realizan un movimiento anterior e inferior de la mandíbula generando variaciones anatómicas en la VAS que consiguen incrementar el área seccional faríngea. Este movimiento estabiliza y fija la mandíbula y el hueso hioides, lo que impide la posterorrotación de estas estructuras durante el decúbito y evita la ocupación de la vía respiratoria.

Se ha comprobado que en los individuos SAHS el porcentaje total de tiempo en el que los pacientes duermen con la boca abierta más de 5 mm (rotación posterior y retrusión) es significativamente mayor (69,3%) que en individuos sanos (11,1%); por tanto, el movimiento de rotación y fijación de la mandíbula durante el sueño puede ser uno de los elementos importantes durante el tratamiento con DAM. Este concepto puede explicar, además, por qué algunos pacientes obtienen buena respuesta con avances pequeños incluso sin que se hayan evidenciado cambios en las VAS durante los estudios de la luz faríngea en vigilia9,10. Aunque su mayor efecto aparece en la zona velofaríngea, tienen repercusiones en todos los segmentos faríngeos. Inicialmente aparece un aumento de rigidez y un ensanchamiento del espacio entre los pilares anterior y posterior de la faringe. Se ha podido corroborar también con resonancia magnética11 que se produce una reducción significativa del espesor de las paredes laterales faríngeas que es superior incluso al incremento luminal anteroposterior12. El palatogloso y el palatofaríngeo se tensan, con lo que se reduce o incluso cesa la vibración de los tejidos blandos13. El paladar blando se desplaza ventralmente y aumenta el calibre de las paredes laterales de la zona velofaríngea11,14, hechos ambos que por sí mismos reducen drásticamente el ronquido15.

La rotación horaria de la mandíbula y el aumento pasivo de la dimensión vertical activan al músculo geniogloso, que será en parte el causante de los cambios que acontecen en la lengua16. Lowe et al17 han comprobado electromiográficamente cómo se incrementa el tono de la musculatura lingual (sobre todo el geniogloso) tras insertar los DAM y cómo este vuelve a descender tras quitar el aparato. La lengua se desplaza hacia delante y ocupa una posición superior, dificultando así su caída hacia atrás durante el sueño18.

Los DAM generan cambios en las presiones faríngeas locales que contribuyen a normalizar las propiedades fisiológicas de la VAS19. La relación anatómica alterada (rotación y avance mandibular) induce un aumento de la estimulación neurosensorial que incrementa el tono muscular motor y reduce la colapsabilidad de la VAS. De la misma forma, se incrementa el vector de acción de los músculos dilatadores faríngeos que, al desplazarse centralmente, aumentan su longitud y mejoran su contractilidad20-21. Los DAM realizan una función similar a la que se produce durante las maniobras de sedación consciente y de resucitación cardiopulmonar en las que, tras asegurar y preservar la permeabilidad de la VAS liberándola de cualquier objeto extraño, se realiza una maniobra de tracción anterior e inferior mandibular22.

El avance funcional mandibular induce cambios en la posición del hueso hioides hacia una posición más adelantada. Aparece ahora una nueva situación de equilibrio de la musculatura suprahioidea que favorecería el aumento de volumen y la permeabilidad de la VAS23. Este incremento de las VAS ha sido documentado por numerosos autores con diferentes sistemas de exploración y técnicas de imagen24-28.

Aunque la respuesta no es la misma en todos los pacientes, algunos autores admiten que hay un efecto dependiente de la dosis29,30, ya que al aumentar el grado de avance mandibular mejora la situación clínica del paciente31-33. Así, conforme incrementamos la protrusión aparece una mejoría tanto en el número como en la gravedad de las desaturaciones.

En las últimas décadas se han utilizado técnicas cefalométricas (CFM) para la evaluación de la VAS y la severidad del SAHS. Una de las mayores deficiencias de la CFM es que sólo permite evaluar la VAS en el plano anteroposterior y no permite identificar los cambios laterales. Estos cambios sí pueden cuantificarse con otros sistemas más sofisticados de imagen, como la tomografía computarizada (TC) y la resonancia magnética (RM) que, aunque dan una información más cuantitativa de la VAS, tienen el inconveniente de estar más sujetos a error que la CFM. Así, el corte tanto en la TC como en la RM debe de ser perfectamente perpendicular a la VAS, puesto que si no se hace de esta forma estaremos midiendo áreas de VAS oblicuas, que serán mayores al calibre luminal real. Otra fuente de errores con estas técnicas deriva de su naturaleza dinámica; es decir, cuando el paciente deglute durante la prueba se reduce falsamente el calibre de la VAS y de nuevo puede dar lugar a interpretaciones erróneas de estenosis de la VAS34,36.

No obstante, a pesar de las carencias señaladas los estudios cefalométricos tienen una serie de ventajas37-39, ya que se aproximan más a una apreciación objetiva de la anatomía real del individuo, es una exploración no invasiva, de bajo coste y prácticamente no necesita colaboración por parte del paciente. Hay numerosos estudios que han documentado diferencias entre los pacientes con y sin SAHS40-49. Entre estas diferencias señalamos algunas de las más significativas: mayor estrechamiento en el área velofaríngea, aumento de dimensiones del paladar blando, distancia hioides-plano mandibular (MPH) aumentada, incremento de los ángulos de compensación craneocervical48,49. El análisis CFM puede complicarse todavía más añadiendo nuevas variables: edad, sexo, biotipo facial, fenotipo, posición del paciente (decúbito o bipedestación).

Otros autores han relacionado algunas características CFM con el fracaso de los DAM44 e incluso con el empeoramiento de la sintomatología que aparece en algunos pacientes tras su colocación50. Esto implica que algunas variaciones esqueléticas, como las que cursan con entrecruzamiento dentario excesivo (sobremordida), pueden llegar a tener dos efectos adversos. Por una parte disminuyen la eficacia del DAM, puesto que se necesita una mayor apertura para conseguir un mínimo avance; en segundo lugar, la posterorrotación necesaria para vencer la sobremordida lleva hacia atrás al hioides y al resto de las estructuras, favoreciendo la reducción de la luz faríngea50-52.

Tangugsorn et al48,53 y Miyamoto et al54 han documentado cambios en la postura de la lengua en posición supina y en bipedestación. En pacientes no apneicos, las posiciones linguales cambian, apareciendo un movimiento dorsal de la lengua hacia la luz faríngea. Este hecho no aparece por el contrario en pacientes SAHS, donde el cambio posicional (durante la vigilia) no implica variaciones en la posición lingual, como un mecanismo protector del colapso aéreo secundario al cambio gravitacional. Esta compensación durante la vigilia está producida por el incremento del tono muscular que los pacientes apneicos tienen respecto de pacientes no SAHS55,56 y que desaparece durante el sueño favoreciendo la obstrucción de la VAS57,58.

También se ha asociado una serie de condiciones esqueléticas y faciales con una respuesta más favorable a la terapia con DAM como son: altura facial anterior reducida, base craneal anterior larga, longitud maxilar aumentada, plano mandibular reducido59, distancia MPH reducida y paladar blando corto60, ausencia de trastornos nasales, grados I y II de Mallampati, sujetos capaces de realizar una protrusión mandibular > 5-7 mm61, proporciones normales entre diámetro de la vía aérea-paladar blando-lengua62. De la misma forma se ha asociado una mejor respuesta a los DAM en pacientes con bajo índice de masa corporal63, circunferencia del cuello reducida, pacientes jóvenes40, índice de apnea-hipopnea (IAH) basal bajo y en episodios apneicos asociados a la posición supina64.

Eficacia de los DAM

Inicialmente se comenzó con la publicación de case reports, luego aparecieron series de casos cada vez más numerosos y ahora, tras varios ensayos clínicos bien controlados13,18,33,65, podemos afirmar que, a pesar de la considerable variabilidad de diseños que encontramos entre los DAM, hay suficientes avales en la bibliografía científica que demuestran su eficacia18,64-66. Todos estos estudios publicados documentan polisomnográficamente la validez del método. Aunque en un principio el diseño del aparato no parecía tener relación con su efectividad67, recientemente se ha descrito que incluso se puede mejorar el porcentaje de éxitos con el uso de los DAM con avance regulable29,30,33, con lo que se concluye que cuanto más agresivos resultan los protocolos, tanto mayor es la relación de éxitos18,68. En este mismo orden de cosas han aparecido algunos estudios69-72 en los que, al igual que se realiza con la CPAP (single-nigth titration), se ha ensayado la titulación de los DAM de avance regulable durante la noche del estudio. Con ello se conseguiría la dosis óptima terapéutica y se acortaría de manera notable el protocolo de avance. Estos ensayos, aunque prometedores, resultan difíciles de estandarizar dada la alta variabilidad de tolerancia y de respuesta individual al tratamiento.

Hay también pacientes en los que no se consiguen buenos resultados: nos referimos a algunos SAHS severos, ya que aunque hay evidencias y datos positivos al respecto, el porcentaje de éxitos resulta inferior en relación con los leves y moderados68,73,74, pacientes con protrusiva deficiente (escaso avance mandibular), gran sobremordida, casos con problemas dentales y/o periodontales y pacientes con patología de la articulación temporomandibular (ATM).

Los DAM son muy eficaces en el tratamiento del ronquido (eliminación en el 50% de los casos)67,73-75 y consiguen una reducción significativa del mismo en 90%-100% de los pacientes65,73,74, así como una mejora en la calidad del sueño67,76,77. El ronquido se asocia también a una entidad intermedia, como es el SARVAS, en el que aparece somnolencia junto con un aumento anormal de la presión esofágica, aunque sin apneas/hipopneas mensurables78. La remisión o reducción del ronquido en estos casos sugiere que los DAM pueden ser el tratamiento de elección79.

En los registros polisomnográficos se confirman los beneficios percibidos por el paciente, con reducción de la frecuencia e intensidad del ronquido, apneas, número de arousals y un aumento significativo del sueño de ondas lentas y REM18,64,80,81.

DAM frente a CPAP

La CPAP es capaz de controlar la mayoría de los casos, mientras que los DAM no poseen esta eficacia. Esta ventaja de base de la CPAP está compensada, en parte, por la mejor aceptación de los DAM por los pacientes. Incluso los individuos con buena respuesta a ambas terapias suelen decantarse hacia los DAM en un porcentaje muy significativo18,33,82. Ferguson et al33 compararon la eficacia, los efectos secundarios, el seguimiento de tratamiento y las preferencias individuales en un estudio prospectivo aleatorizado. El IAH era inferior con CPAP (3,5 ± 1,6) que con DAM (9,7 ± 7,3). Otros estudios18,74 muestran cifras similares en cuanto a la saturación de oxígeno, que también es superior con la CPAP. Pero al analizar la satisfacción personal, el seguimiento y los efectos secundarios, los DAM fueron superiores a la CPAP. La preferencia de los DAM frente a otras terapias se explica por diferentes razones, entre las que destacan la comodidad, el ruido, la necesidad de abastecimiento eléctrico, la claustrofobia y la irritación tanto cutánea como de la VAS. Se ha publicado que el uso de los DAM es menos efectivo que la CPAP en pacientes con SAHS severos y que debe limitarse a casos leves y moderados18,64,82. Otros estudios, sin embargo, no encuentran razones para excluir a algunos de estos pacientes y abogan por su utilización incluso en casos severos, con buenos resultados aunque con gran variabilidad individual65,75,76,83.

Uvulopalatofaringoplastia frente a DAM

En relación con algunos procedimientos quirúrgicos, como la uvulopalatofaringoplastia, hay trabajos donde los DAM se confirman superiores en cuanto a preferencia individual y eficacia de tratamiento84,85. Pero además, incluso en situaciones de fracasos quirúrgicos86, los DAM han sido utilizados con éxito como "tratamiento de rescate" y han mejorado la sintomatología en más del 50% de los pacientes84,87. Algunos aparatos intraorales, como los elevadores del paladar (Palatal Lifting Devices), a pesar de que son poco utilizados por sus escasos resultados y la escasa tolerancia por parte de los pacientes, pueden utilizarse como factores stent tras una uvulopalatofaringoplastia, ya que dificultan la retracción cicatrizal generada por ésta88.

Protocolos de actuación

Al comparar los riesgos y los beneficios del tratamiento con DAM respecto a las otras terapias disponibles podemos sugerir que los DAM representan una alternativa útil para estos pacientes. Los criterios ASDA que antes hemos mencionado suponen un buen punto de partida, pero también hay otros factores que pueden ayudar al clínico de la unidad del sueño a remitir al paciente a un profesional cualificado en el manejo de los DAM.

Sí DAM

­ Ronquido: síntoma principal.

­ Rechazo/intolerancia a la CPAP.

­ Fallo quirúrgico.

­ Viajes frecuentes.

­ Claustrofobia y/o problemas de ajuste con la mascarilla.

No DAM

­ Somnolencia: síntoma principal.

­ Dentición insuficiente para dar estabilidad al DAM.

­ Enfermedad activa de ATM.

­ Facilidad para las naúseas.

­ Obesidad mórbida.

­ Desaturación de O2 severa.

Si el ronquido es el síntoma principal, hay poca sintomatología diurna (somnolencia) y la hipoxemia no es un factor preponderante, los DAM pueden ser el tratamiento de elección. Pero cuando la somnolencia es el principal síntoma, la CPAP supone una mejor elección, puesto que la respuesta positiva al tratamiento mejora y asegura la aceptación de éste18,33. Los pacientes dependientes de CPAP pueden utilizar también los DAM como coadyuvantes del tratamiento, sustituyendo al aparato de presión positiva ante situaciones especiales, como puede ser la realización de un viaje. Los casos severos con hipoxemia y somnolencia diurna, por el contrario, pueden estar infratratados si sólo se utilizan los DAM y estos pacientes se adaptan mejor a la CPAP o a la terapia combinada.

La continuidad de tratamiento con DAM tampoco es perfecta y la mayoría de los datos publicados corresponden a pacientes con buenos resultados. No obstante, la experiencia con CPAP89 sugiere que los cuestionarios realizados a los pacientes quizá tiendan a sobrestimar su actual utilización90. Fergusson et al18,33 detallan hasta un 24% de incumplimiento en pacientes con DAM sin avance y sólo un 5% en los que utilizan DAM regulable. Otros autores objetivan que el seguimiento a largo plazo suele ser bastante regular, con una media de utilización nocturna de 6,8 h/noche68,91 (rango, 5,6-3,5 h).

La caída o expulsión del aparato durante la noche era algo frecuente hace unos años pero ahora, con los nuevos diseños y los nuevos materiales que se han incorporado, ha quedado reducida a alguna anécdota aislada.

Los efectos secundarios pueden diferenciarse en tres tipos: cefalométricos, entre los que aparecen pequeños movimientos mandibulares (avance 0,1 mm; descenso 0,3 mm) rotación mandibular posterior (0,5º), aumento de la longitud mandibular (0,4 mm) aunque este último dato corresponde a los casos con grandes avances. En segundo lugar aparecen pequeños cambios oclusales con reducción de resalte (0,2-1,3 mm) y sobremordida (0,1-1,8 mm), vestibuloversión incisiva superior (2º), linguoversión incisiva inferior (4,5º) y movimiento mesial de los molares inferiores (0,4 mm). En tercer lugar se describen las molestias universales ante el inicio de cualquier tratamiento ortodóncico y que van disminuyendo con el tiempo de uso del aparato: exceso de salivación, sequedad de mucosas, sensibilidad dental, molestias articulares (ATM) y musculares (catalogadas como dolor miofascial que aparece al quitar el DAM y persisten durante un tiempo). La mayoría de autores ha encontrado una mayor incidencia de estos efectos en los DAM de avance fijo respecto de los de avance regulable. Con los protocolos más agresivos de avance quizá debería esperarse una mayor incidencia de estas complicaciones, pero hasta el momento esto no ha sido así76-77,92-96.

La principal razón para dejar el tratamiento es la pérdida de eficacia; sin embargo, este hecho quizá pueda paliarse con la ayuda de los dispositivos de avance regulable, aunque con ambos se han documentado éxitos terapéuticos incluso en obstrucciones faríngeas más bajas97. En cuanto a las contraindicaciones absolutas para la utilización de los DAM, podemos encontrar el insuficiente número de dientes para soportar/retener el aparato, la presencia de una enfermedad activa de la articulación temporomandibular y, finalmente, la movilidad dentaria grave y/o la enfermedad periodontal no tratada98. Pero incluso en pacientes en los que concurran dos de las contraindicaciones absolutas para los DAM (edéntulos totales y enfermedad de la articulación temporomandibular), todavía queda un tipo de aparatología intraoral que se debe tener en cuenta: los aparatos retenedores de la lengua (TRD) que, además, pueden utilizarse con éxito en los pacientes que presentan macroglosia franca y en los que por diferentes razones no pueden realizar un avance mandibular adecuado2,99-102.

El manejo de los DAM requiere formación dental cualificada, tanto para el diseño como para su ajuste y mantenimiento. La fabricación del aparato supone sólo el primer eslabón del tratamiento. Los DAM deben de acomodarse según unos protocolos para alcanzar resultados óptimos, tanto de eficacia como de comodidad para los usuarios. Este proceso puede llevar semanas o incluso meses y debe de evaluarse juntamente con las unidades del sueño. La European Dental Sleep Medicine Academy103 ha realizado un protocolo en detalle para la utilización de estos aparatos. Este mismo organismo recomienda que antes de colocar un DAM en un paciente SAHS se realicen dos cosas: la primera, un diagnóstico y una prescripción del aparato en regla enviado al dentista con formación especializada en este tipo de aparatos, por parte de los médicos responsables de las unidades del sueño, y la segunda, un examen oral adecuado realizado por el profesional encargado del diseño y colocación del DAM (historia médica y dental, valoración de tejidos blandos, evaluación periodontal, examen oclusal y temporomandibular, evaluación de hábitos y parafunciones, examen detallado de los dientes y restauraciones presentes, así como las radiografías y modelos de escayola pertinentes). Es el clínico experimentado quien selecciona y diseña el aparato más indicado para cada caso en particular. Una vez que el aparato ha sido realizado y ajustado se entrena al paciente para su inserción y los cuidados de mantenimiento. Tras conseguir acostumbrar al paciente al aparato se comienza a realizar el avance gradualmente (suele requerir semanas o meses de ajuste) hasta que se consigue una posición confortable y adecuada para aliviar el ronquido y/o la apnea del sueño. Tras esto el paciente debe ser revaluado en la unidad del sueño. Ante cualquier cambio negativo ha de realizarse una nueva visita al laboratorio del sueño y realizar de nuevo el diagnóstico. En estas circunstancias puede llegar a ser preceptivo un posible rediseño o incluso el cambio de aparato. Los pacientes en tratamiento con DAM necesitan una media de visitas regulares durante los primeros 12 meses para asegurar el bienestar, la retención y los resultados adecuados para la enfermedad.

Los DAM son una alternativa razonable pues no suponen un cambio permanente en los individuos (como ocurre con la cirugía) y pueden desestimarse en cualquier momento. El estudio adecuado del paciente en cuanto a evaluación de riesgos/beneficios ha de exponerse individualmente104 para poder conseguir la mejor terapia en cada caso.

Queda todavía un aspecto por plantear que debe tomarse en consideración. Es la posible prevención de la enfermedad que nos ocupa en edades tempranas, en las que con medios ortodóncicos es viable estimular o inhibir el crecimiento sagital y/o transversal mandibular o maxilar y, por ende, inducir cambios en las VAS, consiguiendo una conformación de la vía respiratoria menos susceptible al colapso105-107.

Bibliografía
[1]
Robin P..
Glosoptosis due to atresia and hypotrophy of the mandible..
Am J Dis Child, 48 (1934), pp. 541-7
[2]
Cartwright RD, Samelson CF..
The effects of a nonsurgical treatment for obstructive sleep apnea: the tongue retaining device..
J Am Med Assoc, 248 (1982), pp. 705-9
[3]
Cartwright RD..
Predicting response to the tongue retaining device for sleep apnea syndrome..
Arch Otolaryngol, 111 (1985), pp. 385-8
[4]
Soll BA, George PT..
Treatment of obstructive sleep apnea with a nocturnal airway patency appliance..
N Engl J Med, 313 (1985), pp. 386-7
[5]
Bear SE, Priest JH..
Sleep apnea syndrome. Correction with surgical advancement of the mandible..
J Oral Surg, 38 (1980), pp. 543-9
[6]
Kingshott RN, Jones DR, Taylor DR, et al..
The efficacy of a novel tongue stabilizing device on polisomnographic variables in sleep disordered breathing: a pilot study..
Sleep Breath, 6 (2002), pp. 69-76
[7]
Dental appliances for the treatment of snoring and obstructive sleep apnea. En: Kryger MH, Roth T, Dement WC, editors. Principles and practice of sleep medicine. Philadelphia: Saunders; 1994. p. 722-35.
[8]
American Sleep Disorders Association Standards of Practice Committee..
Practice parameters for the treatment of snoring and obstructive sleep apnea with oral appliances..
Sleep, 18 (1995), pp. 511-3
[9]
Miyamoto K, Ozbek M, Lowe AA, et al..
Mandibular posture during sleep in healthy adults..
Arch Oral Biol, 43 (1998), pp. 269-75
[10]
Miyamoto K Ozbek M, Lowe AA, et al..
Mandibular posture during sleep in patients with obstructive sleep apnea..
Arch Oral Biol, 43 (1999), pp. 657-64
[11]
Schwab RJ, Gupta KB, Duong D, et al..
Upper airway soft tissue structural changes with dental appliances in apneics..
Am J Respir Crit Care Med, 153 (1996), pp. A719
[12]
Ryan CF, Love LL, Peat D, Fleetham JA, Lowe AA..
Mandibular advancement oral appliance therapy for obstructive sleep apnea: effect on awake caliber of the velopharynx..
Thorax, 54 (1999), pp. 9729-77
[13]
Hoekema A, Stegena B, de Bont LG.M, et al..
Efficacy and co-morbidity of oral appliances in the treatment of obstructive sleep apnea-hypopnea:a systematic review Clin Rev Oral Biol Med, 15 (2004), pp. 137-55
[14]
Ryan CF, Lowe AA, Li D, et al..
Three-dimensional upper airway computedtomography in obstructive sleep apnea..
Am Rev Respir Dis, 144 (1991), pp. 428-32
[15]
De Carlos F, Cobo J, Díaz Esnal B, et al..
Ronchopatie chronique et syndrome de l'apnée-hypopnée obstructive du sommeil chez l'enfant..
L'orthodontie française, 74 (2003), pp. 431-57
[16]
Lowe AA, Gionhaku N, Takeuchi K, Fleetham JA..
Three-dimensional CT reconstructions of tongue and airway in adult subjects with obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 90 (1986), pp. 364-74
[17]
Effects of mandibular repositioning appliance used in the treatment of obstructive sleep apnea on muscle activity. En: Suratt PM, Remmers JE, editors. Sleep and respiration. New York: Willey; 1990. p. 395-405.
[18]
Ferguson KA, Ono T, Lowe AA, Al-Majed S, Love LI, Fleetham JA..
A short term controlled trial of an adjustable oral appliance for the treatment of mild to moderate obstructive sleep apnea..
Thorax, 52 (1997), pp. 362-8
[19]
Yoshida K..
Effect of a prosthetic appliance for treatment of sleep apnea syndrome on masticatory and tongue muscle activity..
J Prosthet Dent, 79 (1998), pp. 537-44
[20]
Adachi S, Lowe AA, Tsuchiya M, et al..
Genioglossus muscle activity an inspiratory timing in obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 104 (1993), pp. 138-45
[21]
Lowe AA, Fleetham J, Rya.n, et al..
Effects of a mandibular repositioning appliance used in the treatment of obstructive sleep apnea on tongue muscle activity..
Prog Clin Biol Res, 345 (1990), pp. 395-405
[22]
Tsuiki S, Lowe AA, Almeida FR, et al..
Effects of mandibular advancement on airway curvature and obstructive sleep apnea severity..
Eur Respir J, 23 (2004), pp. 263-8
[23]
Battagel JM, Johal A, L'Estrange PR, Croft CB, Kotecha B..
Changes in airway and hyoid position in response to mandibular protrusion in subjects with obstructive sleep apnoea (OSA)..
Eur J Orthod, 2 (1999), pp. 363-76
[24]
Lowe A, Gionhaku N..
Three dimensional CT reconstruction of the tongue and airway in adults subjects with obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 90 (1986), pp. 364-74
[25]
Cobo J, Canut JA, Carlos F..
Changes in the upper airway of patients who wear a modified functional appliance to treat obstructive sleep apnea..
Int J Adult Orthod Orthognath Surg, 10 (1995), pp. 53-7
[26]
Isono SA, Tanaka A, Sho Y, et al..
Advancement of mandible improves velopharyngeal airway patency..
J Appl Physiol, 79 (1995), pp. 2132-8
[27]
Battagel JM, L'Estrange PR, Nolan P, Harkness B..
The role of lateral cephalometric radiography and fluoroscopy in assessing mandibular advancement in sleep-related disorders..
Eur J Orthod, 20 (1998), pp. 121-32
[28]
Gale DJ, Sawyer RH, Woodcock A, et al..
Do oral appliances enlarge the airway in patients with obstructive sleep apnoea? A prospective computerized tomographic study..
Eur J Orthod, 22 (2000), pp. 159-68
[29]
Kato J, Isono SA, Tanaka A, et al..
Dose-dependent effects of mandibular advancement of pharyngeal mechanics and nocturnal oxygenation in patients with sleep-disordered breathing..
Chest, 117 (2000), pp. 1065-72
[30]
Walker-Engstrom ML, Ringqvist I, Vestling O, et al..
A prospective randomized Study comparing two different degrees of mandibular advancement with a dental appliance in treatment of severe obstructive sleep apnea..
Sleep Breath, 7 (2003), pp. 119-30
[31]
Gao X, Otsuka R, Ono T, et al..
Effect of tirated mandibular advancement and jaw opening on the upper airway in nonapneic men: a magnetic resonance imaging and cephalometric study..
Am J Orthod Dentofac Orthop, 125 (2004), pp. 191-9
[32]
Marklund M, Franklin C, Sahlin C, et al..
The effect of mandibular advancement device on apneas and sleep in patients with OSAS..
Chest, 113 (1998), pp. 707-13
[33]
Ferguson KA, Ono T, Lowe AA, et al..
A randomized crossover study of an oral appliance vs nasal continuous positive airway pressure in the treatment of mild moderate obstructive sleep apnea..
Chest, 109 (1996), pp. 1269-75
[34]
Liu Y, Lowe AA, Fleetham JA, et al..
Cephalometric and physiologic predictors of the efficacy of an adjustable oral appliance for treating obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 120 (2001), pp. 639-47
[35]
Gao XM, Zeng XL, Fu MK, et al..
Magnetic resonance imaging of the upper airway in obstructive sleep apnea before and after oral appliance therapy..
Chin J Dent Res, 2 (1999), pp. 27-35
[36]
Shepard JW, Garrison M, Wentzel V..
Upper airway distensibility and collapsability in patients with obstructive sleep apnea..
Chest, 98 (1990), pp. 84-91
[37]
Skinner MA Robertson CJ, Kingshot RN, et al..
The efficacy of a mandibular advancement splint in relation to cephalometric variables..
Sleep Breath, 6 (2002), pp. 115-24
[38]
Maltais F, Carrier G, Gormier Y, Series F..
Cephalometric measurements in snorers, non snorers and patients with sleep apnea..
Thorax, 46 (1991), pp. 419-23
[39]
Zucconi M, Ferini Strambi L, Palazi S, et al..
Habitual snoring with and without obstructive sleep apnea: the importance of cephalometric variables..
Thorax, 47 (1992), pp. 157-61
[40]
Partinen M, Guilleminault C, Quera Salva MA, et al..
Obstructive sleep apnea and cephalometric roentgenograms. The role of anatomic upper airway abnormalities in the definition of abnormal breathing during sleep..
Chest, 93 (1988), pp. 1199-205
[41]
Lyberg.T, Krogstad O, Djupesland G..
Cephalometric analysis in patients with obstructive sleep apnea syndrome: skleletal morpphology..
J Laryngol Otol, 103 (1989), pp. 287-92
[42]
Tsuchiya M, Lowe AA, Pae EK..
Obstructive sleep apnea subtipes by cluster analysis..
Am J Orthod Dentofanc Orthop, 101 (1992), pp. 542-53
[43]
Chabolle F, Lachiver X, Fleury B..
L' interêt physiopathologique d'une étude cephalometrique par téléradiographie et IRM dans le syndrome d'apnée du sommeil, deductions thérapeutiques..
Ann Otolaryng (Paris), 107 (1990), pp. 156-9
[44]
Lowe AA, Ono T, Fergusson KA, et al..
Cephalometric comparisons of craniofacial and upper airway structure by skeletal subtype and gender in patients with obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 110 (1996), pp. 653-64
[45]
Mayer P, Pepin JL, Bettega G, et al..
Relationship between body max index, age and upper airway measurements in snorers and sleep apnea patients..
Eur J Respir, 9 (1996), pp. 1801-9
[46]
Battagel JM, L'Estrange PR..
The cephalometric morphology of patients with obstructive sleep apnea (OSA)..
Eur J Orthod, 18 (1996), pp. 557-69
[47]
Solow B, Skov S, Ovesen J, et al..
Airway dimensions and head posture in obstructive sleep apnea..
Eur J Orthod, 18 (1996), pp. 571-9
[48]
Tangugsorn V, Skatvedt O, Krogstad O, Lyberg T..
Obstructive sleep apnea: a cephalometric study. Part I. Cervico-craniofacial skeletal morphology..
Eur J Orthod, 17 (1995), pp. 45-56
[49]
Ozbek M, Miyamoto K, Fleetham JA..
Natural head posture, upper airway morphology and obstructive sleep apnea severity in adults..
Eur J Orthod, 20 (1998), pp. 133-43
[50]
De Carlos F, Cobo J, Fernández Mondragón MP, Díaz-Esnal B..
Relación entre la maloclusión e incremento de la vía aérea utilizando un aparato intraoral en pacientes con síndrome de apnea obstructiva del sueño..
RCOE, 7 (2002), pp. 359-64
[51]
Skinner MA Robertson CJ, Kingshot RN, et al..
The efficacy of a mandibular advancement splint in relation to cephalometric variables..
Sleep Breath, 6 (2002), pp. 115-24
[52]
Pitsis A, Darendelier M, Gosoptoulos H..
Effect of vertical dimension on efficacy of oral appliance therapy in obstructive sleep apnea..
Am J Respir Crit Care Med, 166 (2002), pp. 860-4
[53]
Tangugsorn V, Skatvedt O, Krogstad O, Lyberg T..
Obstructive sleep apnea: a cephalometric study. Part II. Uvuloglossopharyngeal morphology..
Eur J Orthod, 15 (1995), pp. 57-67
[54]
Miyamoto K, Ozbek MM, Lowe AA, Fleetham JA..
Effect on body position on tongue posture in awake patients with obstructive sleep apnea..
Thorax, 52 (1997), pp. 255-9
[55]
Mezzanotte WS, Tangel DJ, White DP..
Influence of sleep onset on upper airway muscle activity in apnea patients versus normal controls..
Am J Respir Crit CAre Med, 153 (1996), pp. 1880-7
[56]
Hendricks JC, Petrof BJ, Panckeri K, Pack AI..
Compensatory hyperactivity of an upper airway dilator in bulldogs..
Am Rev Respir Dis, 148 (1993), pp. 185-94
[57]
Schwartz AR, Eisele DW, Hari A, et al..
Electrical stimulation of the lingual musculature in obstructive sleep apnea..
J Appl Physiol, 81 (1996), pp. 643-52
[58]
Petrof BJ Hendricks JC, Pack AI..
Does upper airway muscle injury trigger a vicious cycle in obstructive sleep apnea? A hypotesis..
Sleep, 19 (1996), pp. 465-71
[59]
Skinner MA, Robertson CJ, Kingshot RN, et al..
The efficacy of a mandibular advancement splint in relation to cephalometric variables..
Sleep Breath, 6 (2002), pp. 115-24
[60]
Efficacy of a herbst mandibular advancement device in obstructive sleep apnea. Am J Crit Care Med. 1994149:905-9.
[61]
Liu Y, Lowe AA, Fleetham JA, et al..
Cephalometric and physiologic predictors of the efficacy of an adjustable oral appliance for treating obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 120 (2001), pp. 639-47
[62]
Gao XM, Zeng XL, Fu MK, et al..
Magnetic resonance imaging of the upper airway in obstructive sleep apnea before and after oral appliance theraphy..
Chin J Dent Res, 2 (1999), pp. 27-35
[63]
Mayer G, Meier-Ewert K..
Cephalometric predictors for orthopaedic mandibular avancement in obstructive sleep apnea..
Eur J Orthod, 17 (1995), pp. 35-43
[64]
Marklund M, Franklin C, Sahlin C, et al..
The effect of mandibular advancement device on apneas and sleep in patients with OSAS..
Chest, 113 (1998), pp. 707-13
[65]
Walker-Engstrom ML, Ringqvist I, Vestling O, et al..
A prospective randomized Study comparing two different degrees of mandibular advancement with a dental appliance in treatment of severe obstructive sleep apnea..
Sleep Breath, 7 (2003), pp. 119-30
[66]
Schmidt-Nowara W, Lowe AA, Wiegand L, et al..
Oral appliances for the treatment of snoring and obstructive sleep apnea: a review..
Sleep, 18 (1995), pp. 501-10
[67]
Schmidt-Nowara W, Lowe AA, Wiegand L, et al..
Oral appliances for the treatment of snoring and obstructive sleep apnea: a review..
Sleep, 18 (1995), pp. 501-10
[68]
Lowe A, Sjoeholm T, Ryan C, et al..
The effects of klearway oral appliance on airway size and obstructive sep apnea..
Sleep, 23 (2000), pp. S172-8
[69]
Raphaelson MA, Alpher EJ, Bakker KW, Pelstrom JR..
Oral appliance theraphy for obstructive sleep apnea syndrome: progressive mandibular advancement during polisomnography..
Cranio, 16 (1998), pp. 44-50
[70]
Tsai WH, Vasquez JC, Oshima T, et al..
Remotely controlled mandibular positioner predicts efficacy of oral appliances in sleep apnea..
Am J Respir Crit Care Med, 170 (2004), pp. 366-70
[71]
Petelle V, Vincent G, Gagnadoux F, et al..
One-nigth mandibular advancement titration for obstructive sleep apnea syndrome: a pilot study..
Am J Respir Crit Care Med, 165 (2002), pp. 1150-3
[72]
Fleury B, Rakotonahary D, Petelle.B, et al..
Mandibular advancement titration for obstructive sleep apnea. Optimization of the procedure by comparing clinical and oximetric parameters..
Chest, 125 (2004), pp. 1761-7
[73]
Schmidt-Nowara W, Meade T, Hays M..
Treatment of snoring and obstructive sleep apnea with a dental orthosis..
Chest, 99 (1991), pp. 1378-85
[74]
O'Sullivan RA, Hillman DR Mateljan R, et al..
Mandibular advancement splint: an appliance to treat snoring and obstructive sleep apnea..
Am J Respir Crit Care Med, 151 (1995), pp. 194-8
[75]
Clark GT, Blumenfeld S, Yoffe N..
A cross over study comparing the efficacy of CPAP with MAD on patients with OSA..
Chest, 109 (1996), pp. 1447-83
[76]
Gosoptoulos H, Chen C, Quian J, et al..
Oral appliance theraphy improves symptoms in obstructive sleep apnea: a randomized controled trial..
Am J Respir Crit Care Med, (2002), pp. 166743-8
[77]
Metha A, Quian J, Petocz P..
A randomized controled study of a mandibular advancement splint for obstructive sleep apnea..
Am J Respir Cri Care Med, 163 (2001), pp. 1457-61
[78]
Guilleminault C, Stoohs R, Clerk A..
A cause of excessive daytime sleepiness: the upper airway resistance syndrome..
Chest, 104 (1993), pp. 781-7
[79]
Loube DI, Andrada T, Shanmagum N, Singer M..
Successful treatment of upper airway resistance syndrome with an oral appliance..
Sleep Breathing, 2 (1997), pp. 97-101
[80]
Bloch KE, Iseli A, Zhang JN, et al..
A randomized controled crossover tial of two oral appliances for sleep apnea treatment..
Am J Respir Crit Care Med, 162 (2000), pp. 246-51
[81]
Menn SJ, Loube DI, Morgan TD, et al..
The mandibular repositioning device: role in the treatment of obstructive sleep apnea..
Sleep, 19 (1996), pp. 794-800
[82]
Henke KG, Franzt DE, Kuna ST, et al..
An oral elastic advancement device for obstructive sleep apnea..
Am J Respir Crit Care Med, 161 (2000), pp. 420-5
[83]
Fransson AM.C, Tegelberg A, Leissner L, et al..
Effects of a mandibular protruding device on the sleep of patients with obstructive sleep apnea and snoring problems. A 2 year follow up..
Sleep Breath, 7 (2003), pp. 131-41
[84]
Wilhemson B, Tegelberg A, Walker-Enstrom ML..
A prospective randomized study of dental appliance compared with UPPP in the treatment of OSA..
Acta Otolaryngol Scand, 119 (1999), pp. 503-9
[85]
Larrosa F, Hernández L, Morello A, et al..
Laser assisted uvulopalatoplasty for snoring:does it meet the expectations? Eur Respir J, 24 (2004), pp. 66-70
[86]
Frinkelstein Y, Stein G, Ophir D..
Laser-assisted UPPP for the management of OSA..
Arch Otolaryngol Head Neck Surg, 128 (2002), pp. 429-38
[87]
Millman RP, Rosemberg LC, Carlisle CC..
The efficacy of oral appliances in the treatment of persistent sleep apnea after UPPP..
Chest, 113 (1998), pp. 992-6
[88]
DeAngelo A, Mysliwiec V..
Resolution of severe sleep disordered breathing with a nasopharyngeal obturator in 2 cases of nasopharyngeal stenosis complicating uvulopalatopharyngoplasty..
Sleep Breath, 8 (2004), pp. 49-55
[89]
Chasens ER, Pack AI, Maislin G, et al..
Claustrophobia and adherence to CPAP treatment..
West J Nurs Res, 27 (2005), pp. 307-21
[90]
Kribbs NB, Pack Ai, Kline LR, et al..
Objective measurements of patterns of nasal CPAP use by patients with obstructive sleep apnea..
Am Rev Respir Dis, 147 (1993), pp. 887-95
[91]
Pancer J, Al-Faifi S, Al-Faifi M, et al..
Evaluation of variable mandibular advancement appliance for treatmet of snoring and sleep apnea..
Chest, 116 (1999), pp. 1511-8
[92]
Martinez Gomis J, Willaert Jiménez-Pallarejo E, Nogués Pujo.l, et al..
Efectos adversos del uso de una férula de avance mandibular en el tratamiento de los transtornos respiratorios del sueño. Revisión de la literatura..
Archivos de Odontoestomatología, 20 (2004), pp. 88-100
[93]
Fritsch KM, Iseli A, Russi EW, Bloch KE..
Side effects of mandibular advancement devices for sleep apnea treatment..
Am J Resp Crit Care Med, 164 (2001), pp. 813-8
[94]
Fransson AM.C, Tegelberg A, Svenson BA.H, et al..
Influence of mandibular protruding device on airway passages and dentofacial characteristics in obstructive apnea and snoring..
Am J Dentofac Orthop, 122 (2002), pp. 371-9
[95]
Bondemark L, Lindman R..
Craniomandibular status and function in patients with habitual snoring and obstructive sleep apnea after nocturnal treatment with a mandibular advancement splint: a 2-year follow-up..
Eur J Orthod, 22 (2000), pp. 53-60
[96]
Ringqvist M, Walker-Engstrom ML, Tegelberg A, et al..
Dental and skeletal changes after 4 years of obstructive sleep apnea treatment with a mandibular advancement device: a prospective randomized study..
Am J Dentofac Orthop, 124 (2003), pp. 53-60
[97]
Conaway JR, Scherr SC..
Multidisciplinary management of the airway in a trauma-induced brain injury patient..
Sleep Breath, 8 (2004), pp. 165-70
[98]
Petit FX, Pépin JL, Bettega G, et al..
Mandibular advancement devices. Rate of contraindications in 100 consecutive obstructive sleep apnea patients..
Am J Respir Crit Care Med, 166 (2002), pp. 274-8
[99]
Cartwright R, Stefoski D, Caldarelli D, et al..
Toward a treatment logic for sleep apnea:the place of tongue retaining device..
Behav Res Ther, 26 (1988), pp. 121-6
[100]
Cartwright R Ristanovic R, Díaz F, et al..
A comparative study of treatments for positional sleep apnea..
Sleep, 14 (1991), pp. 546-52
[101]
Ono T, Lowe AA, Ferguson KA, et al..
The effect of tongue retaining device on awake genioglossus muscle activity in patients with obstructive sleep apnea..
Am J Orthod Dentofac Orthop, 110 (1996), pp. 28-35
[102]
Ono T, Lowe AA, Ferguson KA, et al..
A tongue retaining device and sleep state genioglossus muscle activity in patients with obstructive sleep apnea..
[103]
Marklund M, Netzer NC..
Report from the first conference of European Dental Sleep Medicine Academy..
Sleep Breath, 9 (2005), pp. 41-2
[104]
Cost justification for diagnosis and treatment of obstructive sleep apnea..
Position statement of the American Academy of Sleep Medicine..
Sleep, 33 (2000), pp. 1017-8
[105]
Cobo J..
El activador y las vías aéreas superiores..
Rev Esp Orthod, 24 (1994), pp. 311-8
[106]
Cobo J, Díaz-Esnal B, De Carlos F..
Ortodoncia y vías aéreas superiores RCOE, 7 (2002), pp. 417-27
[107]
Cobo J, De Carlos F, Macias E..
L'Orthodontie et les voies aériennes supérieures..
L'Orthodontie Française, 75 (2004), pp. 31-7
Archivos de Bronconeumología
Article options
Tools

Are you a health professional able to prescribe or dispense drugs?