array:23 [
  "pii" => "S0300289615324728"
  "issn" => "03002896"
  "doi" => "10.1016/S0300-2896(15)32472-8"
  "estado" => "S300"
  "fechaPublicacion" => "1980-07-01"
  "aid" => "72541"
  "copyright" => "Sociedad Española de Neumología y Cirugía Torácica"
  "copyrightAnyo" => "1980"
  "documento" => "article"
  "crossmark" => 0
  "subdocumento" => "fla"
  "cita" => "Arch Bronconeumol. 1980;16:111-6"
  "abierto" => array:3 [
    "ES" => true
    "ES2" => true
    "LATM" => true
  ]
  "gratuito" => true
  "lecturas" => array:2 [
    "total" => 1183
    "formatos" => array:3 [
      "EPUB" => 41
      "HTML" => 693
      "PDF" => 449
    ]
  ]
  "itemSiguiente" => array:18 [
    "pii" => "S030028961532473X"
    "issn" => "03002896"
    "doi" => "10.1016/S0300-2896(15)32473-X"
    "estado" => "S300"
    "fechaPublicacion" => "1980-07-01"
    "aid" => "72542"
    "copyright" => "Sociedad Española de Neumología y Cirugía Torácica"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Arch Bronconeumol. 1980;16:117-9"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 1864
      "formatos" => array:3 [
        "EPUB" => 63
        "HTML" => 779
        "PDF" => 1022
      ]
    ]
    "es" => array:10 [
      "idiomaDefecto" => true
      "titulo" => "Edema pulmonar neurogenico. Descripcion de un caso"
      "tienePdf" => "es"
      "tieneTextoCompleto" => 0
      "tieneResumen" => array:2 [
        0 => "es"
        1 => "en"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "117"
          "paginaFinal" => "119"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Neurogenic pulmonary edema. Description of one case"
        ]
      ]
      "contieneResumen" => array:2 [
        "es" => true
        "en" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "H. Fité Reig, J. Oriol Sabater, J. Ruiz Manzano, C. Plans Bolívar, M. Vilaplana Soler"
          "autores" => array:5 [
            0 => array:2 [
              "nombre" => "H."
              "apellidos" => "Fité Reig"
            ]
            1 => array:2 [
              "nombre" => "J."
              "apellidos" => "Oriol Sabater"
            ]
            2 => array:2 [
              "nombre" => "J."
              "apellidos" => "Ruiz Manzano"
            ]
            3 => array:2 [
              "nombre" => "C."
              "apellidos" => "Plans Bolívar"
            ]
            4 => array:2 [
              "nombre" => "M."
              "apellidos" => "Vilaplana Soler"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S030028961532473X?idApp=UINPBA00003Z"
    "url" => "/03002896/0000001600000003/v1_201509151256/S030028961532473X/v1_201509151256/es/main.assets"
  ]
  "itemAnterior" => array:18 [
    "pii" => "S0300289615324716"
    "issn" => "03002896"
    "doi" => "10.1016/S0300-2896(15)32471-6"
    "estado" => "S300"
    "fechaPublicacion" => "1980-07-01"
    "aid" => "72540"
    "copyright" => "Sociedad Española de Neumología y Cirugía Torácica"
    "documento" => "article"
    "crossmark" => 0
    "subdocumento" => "fla"
    "cita" => "Arch Bronconeumol. 1980;16:106-10"
    "abierto" => array:3 [
      "ES" => true
      "ES2" => true
      "LATM" => true
    ]
    "gratuito" => true
    "lecturas" => array:2 [
      "total" => 1778
      "formatos" => array:3 [
        "EPUB" => 71
        "HTML" => 698
        "PDF" => 1009
      ]
    ]
    "es" => array:10 [
      "idiomaDefecto" => true
      "titulo" => "Pleuritis tuberculosa. Diagnostico histologico y bacteriologico en 310 pacientes"
      "tienePdf" => "es"
      "tieneTextoCompleto" => 0
      "tieneResumen" => array:2 [
        0 => "es"
        1 => "en"
      ]
      "paginas" => array:1 [
        0 => array:2 [
          "paginaInicial" => "106"
          "paginaFinal" => "110"
        ]
      ]
      "titulosAlternativos" => array:1 [
        "en" => array:1 [
          "titulo" => "Tuberculous pleurisy. Histological and bacteriological diagnoses in 310 patients"
        ]
      ]
      "contieneResumen" => array:2 [
        "es" => true
        "en" => true
      ]
      "contienePdf" => array:1 [
        "es" => true
      ]
      "autores" => array:1 [
        0 => array:2 [
          "autoresLista" => "C. Plans Bolíbar, A. Aranda Torres, A. Roca Montanari, R. Vidal Pla, S. Sendra Salillas, M. Vilaplana Soler, F. Morell Brotad, J. Morera Prat"
          "autores" => array:8 [
            0 => array:2 [
              "nombre" => "C."
              "apellidos" => "Plans Bolíbar"
            ]
            1 => array:2 [
              "nombre" => "A."
              "apellidos" => "Aranda Torres"
            ]
            2 => array:2 [
              "nombre" => "A."
              "apellidos" => "Roca Montanari"
            ]
            3 => array:2 [
              "nombre" => "R."
              "apellidos" => "Vidal Pla"
            ]
            4 => array:2 [
              "nombre" => "S."
              "apellidos" => "Sendra Salillas"
            ]
            5 => array:2 [
              "nombre" => "M."
              "apellidos" => "Vilaplana Soler"
            ]
            6 => array:2 [
              "nombre" => "F."
              "apellidos" => "Morell Brotad"
            ]
            7 => array:2 [
              "nombre" => "J."
              "apellidos" => "Morera Prat"
            ]
          ]
        ]
      ]
    ]
    "idiomaDefecto" => "es"
    "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0300289615324716?idApp=UINPBA00003Z"
    "url" => "/03002896/0000001600000003/v1_201509151256/S0300289615324716/v1_201509151256/es/main.assets"
  ]
  "es" => array:12 [
    "idiomaDefecto" => true
    "titulo" => "Curvas de flujo-volumen (MEFV). Registro grafico versus osciloscopico"
    "tieneTextoCompleto" => 0
    "paginas" => array:1 [
      0 => array:2 [
        "paginaInicial" => "111"
        "paginaFinal" => "116"
      ]
    ]
    "autores" => array:2 [
      0 => array:3 [
        "autoresLista" => "F. Burgos, R. Rodríguez Roisin, F. Añanos, C. Gistau, A. Agustí Vidal"
        "autores" => array:5 [
          0 => array:2 [
            "nombre" => "F."
            "apellidos" => "Burgos"
          ]
          1 => array:2 [
            "nombre" => "R."
            "apellidos" => "Rodríguez Roisin"
          ]
          2 => array:2 [
            "nombre" => "F."
            "apellidos" => "Añanos"
          ]
          3 => array:2 [
            "nombre" => "C."
            "apellidos" => "Gistau"
          ]
          4 => array:2 [
            "nombre" => "A."
            "apellidos" => "Agustí Vidal"
          ]
        ]
        "afiliaciones" => array:1 [
          0 => array:2 [
            "entidad" => "Servicio de Neumología y Alergia Respiratoria. Hospital Clínico y Provincial. Departamento de Física Médica. Facilitad de Medicina. Barcelona"
            "identificador" => "aff0005"
          ]
        ]
      ]
      1 => array:3 [
        "autoresLista" => "M.M. Rotger, D. Navajas"
        "autores" => array:2 [
          0 => array:3 [
            "nombre" => "M.M."
            "apellidos" => "Rotger"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">&#42;</span>"
                "identificador" => "aff0010"
              ]
            ]
          ]
          1 => array:3 [
            "nombre" => "D&#46;"
            "apellidos" => "Navajas"
            "referencia" => array:1 [
              0 => array:2 [
                "etiqueta" => "<span class="elsevierStyleSup">&#42;</span>"
                "identificador" => "aff0010"
              ]
            ]
          ]
        ]
        "afiliaciones" => array:1 [
          0 => array:3 [
            "entidad" => "Departamento de F&#237;sica M&#233;dica&#46; Servicio de Neumolog&#237;a y Alergia Respiratoria&#46; Hospital Cl&#237;nico y Provincial&#46; Departamento de F&#237;sica M&#233;dica&#46; Facilitad de Medicina&#46; Barcelona"
            "etiqueta" => "&#42;"
            "identificador" => "aff0010"
          ]
        ]
      ]
    ]
    "titulosAlternativos" => array:1 [
      "en" => array:1 [
        "titulo" => "Flow-volume curves &#40;MEFV&#41;&#46; Graphic versus oscilloscopic register"
      ]
    ]
    "pdfFichero" => "main.pdf"
    "tienePdf" => true
    "fechaRecibido" => "1980-05-13"
    "tieneResumen" => true
    "resumen" => array:2 [
      "es" => array:2 [
        "titulo" => "Resumen"
        "resumen" => "<span id="absec0005" class="elsevierStyleSection elsevierViewall"><p id="sp0005" class="elsevierStyleSimplePara elsevierViewall">Para averiguar la fiabilidad del registro de las curvas flujo-volumen &#40;MEFV&#41; con un registrador gr&#225;fico X-Y &#40;H&#46;P&#46;&#44; 7045 A&#41;&#44; se determin&#243; experimentalmente <span class="elsevierStyleItalic">su respuesta en frecuencias y la composici&#243;n espectral de las curvas flujo-tiempo</span> &#40;V-t&#41;&#46; La falta de linealidad del registrador no permiti&#243;&#44; a la vista de los resultados obtenidos&#44; asegurar la absoluta fiabilidad del registro de la curva de MEFV&#46; El objetivo del presente trabajo consisti&#243; en cotejar los registros gr&#225;ficos con los oscilosc&#243;picos &#40;que no ofrecen tales problemas t&#233;cnicos&#41;&#44; simult&#225;neamente recogidos&#44; y evaluar las correlaciones obtenidas entre los valores medidos con cada uno de ellos&#46; Seis individuos&#44; en situaciones cl&#237;nicas distintas&#44; fueron explorados espirom&#233;tricamente&#44; recogi&#233;ndose un total de 23 curvas de MEFNL Los flujos a&#233;reos m&#225;ximos &#40;PEF&#44; V<span class="elsevierStyleInf">25&#37;</span>&#44; etc&#8230;&#41; fueron obtenidos a trav&#233;s del registrador gr&#225;fico &#40;ya se&#241;alado&#41;&#44; y de un osciloscopio &#40;5113&#44; con amplificador 5A15N&#44; Tektronix&#41; acoplados en paralelo y medidos separada e independiente por 3 observadores con experiencia&#46; Ninguno de los registros super&#243; los 10-12 cm&#46; de altura&#46;</p><p id="sp0010" class="elsevierStyleSimplePara elsevierViewall">Los coeficientes de correlaci&#243;n intraindividuales &#40;r&#41; entre ambos registros fueron superiores en su mayor&#237;a a 0&#44;95 &#40;p &#60; 0&#44;001&#41; para los flujos a&#233;reos m&#225;ximos y la capacidad vital forzada&#59; los interindividuales&#44; fueron superiores a 0&#44;99 &#40;p &#60; 0&#44;001&#41; para los mismos par&#225;metros y para ambos registros&#46; Los 3 observadores coincidieron en apreciar la mayor facilidad de medici&#243;n gr&#225;fica&#46; Se concluye que este modelo de registro gr&#225;fico de las curvas de MEFV re&#250;ne las condiciones &#243;ptimas desde un punto de vista t&#233;cnico&#44; siempre y cuando su formato no supere los 10-12 cm de altura&#46;</p></span>"
      ]
      "en" => array:2 [
        "titulo" => "Summary"
        "resumen" => "<span id="absec0010" class="elsevierStyleSection elsevierViewall"><p id="sp0015" class="elsevierStyleSimplePara elsevierViewall">In order to ascertain the reliability of the register of the flow-volume curves &#40;MEFV&#41; with a graphic register X-Y &#40;H&#46;P&#46;&#44; 7045 A&#41;&#44; its <span class="elsevierStyleItalic">response in frequencies and the spectral composition of the flow-time curves</span> &#40;V-t&#41; were determined experimentally&#46; The lack of linearity of the register&#44; given the results obtained&#44; did not permit the authors to assure the absolute reliability of the register of the curve of MEFV&#46; The objective of the present work consisted in comparing the graphic registers with oscilloscopic registers &#40;that do not offer said technical problems&#41;&#44; gathered simultaneously&#44; and evaluating the correlations obtained between the average val&#250;es with each of them&#46; Six individu&#225;is&#44; in different clinical situations&#44; were explored spirometrically&#44; gathering a total of 23 curves of MEFV&#46; The m&#225;ximum aerial flows &#40;PEF&#44; V<span class="elsevierStyleInf">25&#37;</span>&#44; etc&#8230;&#41; were obtained through the graphic register &#40;already indicated&#41;&#44; and from an oscilloscope &#40;5113&#44; with 5A15N amplifier&#44; Tektronix&#41; connected in parallel and measured separtely and independently by 3 observers with experience&#46; None of the registers surpassed 10-12 cm of height&#46;</p><p id="sp0020" class="elsevierStyleSimplePara elsevierViewall">The intraindividual coefficients of correlation &#40;r&#41; between both registers were superior&#44; in the majority of cases&#44; to 0&#46;95 &#40;p &#60; 0&#46;001&#41; for the m&#225;ximum aerial flows and the forced vital capacity&#59; the interindividuals&#44; were superior to 0&#46;99 &#40;p &#60; 0&#46;001&#41; for the same parameters and for both registers&#46; The three observers coincided in noting the greater casiness of graphic measurement&#46; The authors conclude that this model of graphic register of the MEFV curves has optimum conditions from the technical viewpoint when its format does not surpass 10-12 cm of height&#46;</p></span>"
      ]
    ]
    "bibliografia" => array:2 [
      "titulo" => "Bibliografia"
      "seccion" => array:1 [
        0 => array:2 [
          "identificador" => "bibs0005"
          "bibliografiaReferencia" => array:8 [
            0 => array:3 [
              "identificador" => "bib0005"
              "etiqueta" => "1&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Basic equipment and methods"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "J&#46;E&#46; Cotes"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Libro" => array:7 [
                        "edicion" => "4&#46;a edici&#243;n"
                        "titulo" => "&#171;Lung Function&#46; Assessment and application in medicine&#187;"
                        "fecha" => "1978"
                        "paginaInicial" => "19"
                        "paginaFinal" => "56"
                        "editorial" => "Blackwell Scientific Publ"
                        "editorialLocalizacion" => "Oxford"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            1 => array:3 [
              "identificador" => "bib0010"
              "etiqueta" => "2&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Measurements of the respiratory system"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "F&#46;P&#46; Primiano Jr&#46;"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "LibroEditado" => array:4 [
                        "titulo" => "&#171;Medical Instrumentation&#46; Application and design&#187;&#46;"
                        "paginaInicial" => "434"
                        "paginaFinal" => "510"
                        "serieFecha" => "1978"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            2 => array:3 [
              "identificador" => "bib0015"
              "etiqueta" => "3&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Etude de la r&#233;ponse en fr&#233;quence des pneumotachographs"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:3 [
                            0 => "R&#46; Peslin"
                            1 => "J&#46; Morinet-Lambert"
                            2 => "C&#46; Duvivier"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Revista" => array:4 [
                        "tituloSerie" => "Bull&#46; Physio-path&#46; resp&#46;"
                        "fecha" => "1972"
                        "volumen" => "8"
                        "paginaInicial" => "1363"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            3 => array:3 [
              "identificador" => "bib0020"
              "etiqueta" => "4&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "The Readout"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "J&#46; Dubovy"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Libro" => array:6 [
                        "titulo" => "&#171;Introduction to Biomedical Electronics&#187;"
                        "fecha" => "1978"
                        "paginaInicial" => "145"
                        "paginaFinal" => "173"
                        "editorial" => "McGraw-Hill&#44; Inc"
                        "editorialLocalizacion" => "New York"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            4 => array:3 [
              "identificador" => "bib0025"
              "etiqueta" => "5&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Graphic Recorders"
                      "autores" => array:1 [
                        0 => array:2 [
                          "colaboracion" => "Tektronix&#44; Inc"
                          "etal" => false
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Libro" => array:6 [
                        "titulo" => "&#171;Biphysics&#187;"
                        "fecha" => "1970"
                        "paginaInicial" => "399"
                        "paginaFinal" => "412"
                        "editorial" => "Tektronic&#44; Inc"
                        "editorialLocalizacion" => "Beaverton&#46; Oregon"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            5 => array:3 [
              "identificador" => "bib0030"
              "etiqueta" => "6&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Limitations of Instruments and Pitfalls in Use"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "R&#46; Cooper"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "LibroEditado" => array:4 [
                        "titulo" => "&#171;Medical Engineering&#187;&#44;"
                        "paginaInicial" => "1010"
                        "paginaFinal" => "1029"
                        "serieFecha" => "1974"
                      ]
                    ]
                  ]
                ]
              ]
            ]
            6 => array:3 [
              "identificador" => "bib0035"
              "etiqueta" => "7&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "Improved technique for recording respiratory data"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:4 [
                            0 => "R&#46; Pollock"
                            1 => "N&#46;L&#46; Gregory"
                            2 => "A&#46; Shaw"
                            3 => "K&#46; Patel"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:2 [
                      "doi" => "10.1016/j.ultrasmedbio.2015.07.024"
                      "Revista" => array:5 [
                        "tituloSerie" => "Med&#46; &#8217;Biol&#46; Eng&#46; &#38; Comput"
                        "fecha" => "1977"
                        "volumen" => "15"
                        "paginaInicial" => "541"
                        "link" => array:1 [
                          0 => array:2 [
                            "url" => "https://www.ncbi.nlm.nih.gov/pubmed/26361999"
                            "web" => "Medline"
                          ]
                        ]
                      ]
                    ]
                  ]
                ]
              ]
            ]
            7 => array:3 [
              "identificador" => "bib0040"
              "etiqueta" => "8&#46;"
              "referencia" => array:1 [
                0 => array:2 [
                  "contribucion" => array:1 [
                    0 => array:2 [
                      "titulo" => "The Fast Fourier Transform"
                      "autores" => array:1 [
                        0 => array:2 [
                          "etal" => false
                          "autores" => array:1 [
                            0 => "E&#46; Oran Brigham"
                          ]
                        ]
                      ]
                    ]
                  ]
                  "host" => array:1 [
                    0 => array:1 [
                      "Libro" => array:4 [
                        "fecha" => "1974"
                        "paginaInicial" => "49"
                        "editorial" => "Prentice-Hall&#44; Inc"
                        "editorialLocalizacion" => "Englewood Cliffs&#46; New Jersey"
                      ]
                    ]
                  ]
                ]
              ]
            ]
          ]
        ]
      ]
    ]
  ]
  "idiomaDefecto" => "es"
  "url" => "/03002896/0000001600000003/v1_201509151256/S0300289615324728/v1_201509151256/es/main.assets"
  "Apartado" => array:4 [
    "identificador" => "9277"
    "tipo" => "SECCION"
    "en" => array:2 [
      "titulo" => "Originales"
      "idiomaDefecto" => true
    ]
    "idiomaDefecto" => "en"
  ]
  "PDF" => "https://static.elsevier.es/multimedia/03002896/0000001600000003/v1_201509151256/S0300289615324728/v1_201509151256/es/main.pdf?idApp=UINPBA00003Z&text.app=https://archbronconeumol.org/"
  "EPUB" => "https://multimedia.elsevier.es/PublicationsMultimediaV1/item/epub/S0300289615324728?idApp=UINPBA00003Z"
]
Share
Journal Information
Vol. 16. Issue 3.
Pages 111-116 (July - September 1980)
Share
Share
Download PDF
More article options
Vol. 16. Issue 3.
Pages 111-116 (July - September 1980)
Full text access
Curvas de flujo-volumen (MEFV). Registro grafico versus osciloscopico
Flow-volume curves (MEFV). Graphic versus oscilloscopic register
Visits
3692
F. Burgos, R. Rodríguez Roisin, F. Añanos, C. Gistau, A. Agustí Vidal
Servicio de Neumología y Alergia Respiratoria. Hospital Clínico y Provincial. Departamento de Física Médica. Facilitad de Medicina. Barcelona
M.M. Rotger*, D. Navajas*
* Departamento de Física Médica. Servicio de Neumología y Alergia Respiratoria. Hospital Clínico y Provincial. Departamento de Física Médica. Facilitad de Medicina. Barcelona
This item has received
Article information
Resumen

Para averiguar la fiabilidad del registro de las curvas flujo-volumen (MEFV) con un registrador gráfico X-Y (H.P., 7045 A), se determinó experimentalmente su respuesta en frecuencias y la composición espectral de las curvas flujo-tiempo (V-t). La falta de linealidad del registrador no permitió, a la vista de los resultados obtenidos, asegurar la absoluta fiabilidad del registro de la curva de MEFV. El objetivo del presente trabajo consistió en cotejar los registros gráficos con los osciloscópicos (que no ofrecen tales problemas técnicos), simultáneamente recogidos, y evaluar las correlaciones obtenidas entre los valores medidos con cada uno de ellos. Seis individuos, en situaciones clínicas distintas, fueron explorados espirométricamente, recogiéndose un total de 23 curvas de MEFNL Los flujos aéreos máximos (PEF, V25%, etc…) fueron obtenidos a través del registrador gráfico (ya señalado), y de un osciloscopio (5113, con amplificador 5A15N, Tektronix) acoplados en paralelo y medidos separada e independiente por 3 observadores con experiencia. Ninguno de los registros superó los 10-12 cm. de altura.

Los coeficientes de correlación intraindividuales (r) entre ambos registros fueron superiores en su mayoría a 0,95 (p < 0,001) para los flujos aéreos máximos y la capacidad vital forzada; los interindividuales, fueron superiores a 0,99 (p < 0,001) para los mismos parámetros y para ambos registros. Los 3 observadores coincidieron en apreciar la mayor facilidad de medición gráfica. Se concluye que este modelo de registro gráfico de las curvas de MEFV reúne las condiciones óptimas desde un punto de vista técnico, siempre y cuando su formato no supere los 10-12 cm de altura.

Summary

In order to ascertain the reliability of the register of the flow-volume curves (MEFV) with a graphic register X-Y (H.P., 7045 A), its response in frequencies and the spectral composition of the flow-time curves (V-t) were determined experimentally. The lack of linearity of the register, given the results obtained, did not permit the authors to assure the absolute reliability of the register of the curve of MEFV. The objective of the present work consisted in comparing the graphic registers with oscilloscopic registers (that do not offer said technical problems), gathered simultaneously, and evaluating the correlations obtained between the average valúes with each of them. Six individuáis, in different clinical situations, were explored spirometrically, gathering a total of 23 curves of MEFV. The máximum aerial flows (PEF, V25%, etc…) were obtained through the graphic register (already indicated), and from an oscilloscope (5113, with 5A15N amplifier, Tektronix) connected in parallel and measured separtely and independently by 3 observers with experience. None of the registers surpassed 10-12 cm of height.

The intraindividual coefficients of correlation (r) between both registers were superior, in the majority of cases, to 0.95 (p < 0.001) for the máximum aerial flows and the forced vital capacity; the interindividuals, were superior to 0.99 (p < 0.001) for the same parameters and for both registers. The three observers coincided in noting the greater casiness of graphic measurement. The authors conclude that this model of graphic register of the MEFV curves has optimum conditions from the technical viewpoint when its format does not surpass 10-12 cm of height.

Full text is only aviable in PDF
Bibliografia
[1.]
J.E. Cotes.
Basic equipment and methods.
«Lung Function. Assessment and application in medicine», 4.a edición, Blackwell Scientific Publ, (1978), pp. 19-56
[2.]
F.P. Primiano Jr..
Measurements of the respiratory system.
«Medical Instrumentation. Application and design»., pp. 434-510
[3.]
R. Peslin, J. Morinet-Lambert, C. Duvivier.
Etude de la réponse en fréquence des pneumotachographs.
Bull. Physio-path. resp., 8 (1972), pp. 1363
[4.]
J. Dubovy.
The Readout.
«Introduction to Biomedical Electronics», McGraw-Hill, Inc, (1978), pp. 145-173
[5.]
Tektronix, Inc.
Graphic Recorders.
«Biphysics», Tektronic, Inc, (1970), pp. 399-412
[6.]
R. Cooper.
Limitations of Instruments and Pitfalls in Use.
«Medical Engineering»,, pp. 1010-1029
[7.]
R. Pollock, N.L. Gregory, A. Shaw, K. Patel.
Improved technique for recording respiratory data.
Med. ’Biol. Eng. & Comput, 15 (1977), pp. 541
[8.]
E. Oran Brigham.
The Fast Fourier Transform.
Prentice-Hall, Inc, (1974), pp. 49
Copyright © 1980. Sociedad Española de Neumología y Cirugía Torácica
Archivos de Bronconeumología
Article options
Tools

Are you a health professional able to prescribe or dispense drugs?